1. Measurement:
1). Principle: cone-plate method (for whole blood).
cone-plate or capillary method (for plasma).
2). Method:
Cone-plate: complete measurement range, pointwise, prompt, steady state method.
Capillary: micro capillary prompt method (pressure sensor).
3). Signal collection technology: High-precision raster subdivision technology.
4). Working mode: simultaneous working with dual-cap piercing probe (with liquid level sensor function), dual-sample plate, dual-methodologies, three testing modules can work simultaneously.
5). Cap-Piercing function: sample cap-piercing probe module for capped sample tube.
2. Working environment:
1). Operating voltage: 100~240 VAC, 50~60 Hz.
2). Power consumption: 350 VA.
3). Operating temperature: 10~30 °C.
4). Humidity: 30~75%.
3. Working parameters:
1). Accuracy: Newtonian fluid <±1%. Non Newtonian fluid <±2%.
2). CV: Newtonian fluid ≤1%. Non Newtonian fluid ≤2%.
3). Throughput: ≤30 s/sample (whole blood). ≤0.5 s/sample (plasma).
4). Shear rate range: (1~200) S-1.
5). Viscosity range: (0~60) mPa·s.
6). Shear force range: (0~12000) mPa.
7). Sampling volume: 200~800