This is a simultaneous oxygen, nitrogen & hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user's requests.
The oxygen is measured as carbon monoxide and carbon dioxide by two non-dispersive infrared detectors, the nitrogen by a thermal conductivity detector and the hydrogen with non dispersive Infra Red detector as H2O.
Principle and Gas flow diagram
The sample is placed in a graphite crucible and the crucible is maintained between the upper and the lower electrode of the impulse furnace. A high current power passes through the crucible to create a high temperature.
The oxide in the sample reacts with the graphite crucible, and is extracted as carbon monoxide (CO), and carried together with the carrier gas. The extracted gas is directly analysed after the dust filter. According to the oxygen concentration, it is determined by CO or CO2 after pass throw oxidizer (Copper Oxide) with a non dispersive Infra-Red analyser (NDIR).
The Hydrogen is determined with a non dispersive Infra Red analyser as H2O after H2 passes through the oxidizer (Copper Oxide).
The Nitrogen in the sample is extracted as nitrogen gas (N2) and its concentration is determined by a thermal conductivity detector (TCD).
Standard
EMGA fulfils requirements of the standard method for analysis of steel titanium, ceramics etc…
EX)
*ISO10720, ISO17053
*JIS G1228
*ASTM E1019, ASTM E1569, ASTM E1409
User-friendly Software
Measurement Window
Simple software allows easy handling.