Designed for bedside use in the hospital or home, the Aridyne air compressor system is a portable
replacement or supplement to a piped medical air system.
Aridyne air compressors are hospital-grade systems that generate clean dry air without the need for an
external condensation system. Air is delivered through DISS air check valves at 50 psi, with flow rates
exceeding 42 lpm. A unique design consisting of chambers, baffles, and a suspended pump ensures that
the Aridyne is one of the quietest air compressors on the market.
The system works by drawing and filtering ambient air through the air intake. The air is compressed to
80 psi and then cooled via the internal cooling coils which condense the moisture from the compressed air.
The moisture is collected in the condensation trap and automatically drained into the evaporation tray.
The cool, dry air is then regulated to 50 psi and delivered to medical devices through multiple outlets.
Key Features
Internal condensation trap removes moisture from the air to ensure delivered air is dry
Engineered to be one of the quietest medical grade air compressors
Suitable for continuous use 24 hours a day, 7 days a week
Aridyne 2000 models have audible and visual high temperature and low pressure alarms,
with DC-powered 6 volt backup
Circuit breaker, elapsed time indicator, and thermal overload protector included
Air pump inlet filter cleans air and muffles piston noise
Provides 42 lpm
continuous flow through
2 DISS outlets.