AmcareMed LED Oxygen Automatic Manifold is mainly suitable for continuous consumption of bottled oxygen hospital.
In use, the number of or even dozens of oxygen cylinders are centrally connected to the manifold pipe. The gas is filtered, decompression and pressure stabilized, and transported to the use site. AmcareMed Medical Automatic Gas Manifold is fully automatic and has pressure regulators with LED Display. Pressure unit provides a bar, Mpa or psi for measurement conversion.
The fully automatic manifold system monitors cylinder tank pressure electronically, automatically changes over to the secondary tank when the primary cylinder tank is depleted and eliminates the need to manually set a priority side. The solenoid valve technology allows us to eliminate the shuttle valve and it can be realized automatically switch do not need operator debugging. In the electrical system power state, it can maintain continuous supply medical gas.
Our automatic gas manifold is divided according to the number of cylinders and configuration, with a variety of structural forms, 2 * 1 cylinder group, 2 * 2 cylinder group, 2 * 3 cylinder group, 2 * 4 cylinder group, 2 * 5 cylinder group for selection, but also according to user needs and environmental requirements for special configuration.
Specifications
* Closed box without interference
* Stainless steel diaphragm reducer, better to control flow
* With alarm system and remote alarming
* Open design to meet requirements for expansion in future
* Sustainable supply and control circuit when the power failure
* Input: 15Mpa, output: 200m3/h,0.4-0.8Mpa
* Input/output signal: 4–20mA, 485bus communication
* Wall or floor mounts available