The MaxVenturi has been specifically designed to work with high-flow humidification systems to deliver CPAP and HFOT to adults. Using the well-established Venturi principle, the MaxVenturi® to mix ambient air and oxygen without the need for piped medical grade air.
The ability to mix ambient air and oxygen without the need for piped medical grade air
All-in-one oxygen/air mixer, MaxO₂ + analyser and flow meter
Long battery life on analyser (approximately 5,000 hours)
Independently adjustable O₂% and flow
Built-in pole clamp
Product Specification
• O₂ measurement Range: 0 to 100%
• Display Resolution: 0.1%
• Accuracy and Linearity: 1% of full scale at constant temperature, relative humidity, and pressure when calibrated at full scale
• Total Accuracy: ±3% actual oxygen level over full operating temperature range
• Response Time: 90% of final value in approximately 15 seconds at 23 ̊C
• No warm-up time required
• Operating Temperature: 15 ̊C to 40 ̊C (59 ̊F to 104 ̊F)
• Storage Temperature: -15 ̊C to 50 ̊C (5 ̊F to 122 ̊F)
• Atmospheric Pressure: 800 to 1013 mBars
• Humidity: 0 to 95% (non-condensing)
• Power Requirements: Two (2) AA alkaline batteries (2 x 1.5 volts)