Product descriptionCooling incubator with an extended temperature range combined with light module ICH 260/470 (without light-dose control) or ICH-LQC 260/470 (with light dose control). Designed for laboratory and research applications requiring precise temperature control and optional illumination.
Important characteristics- Temperature range: +5 °C to +60 °C
- Proven APT.line™ preheating chamber technology
- Energy-efficient cooling with modern inverter cooling unit
- Cooling with climate-neutral R-290 refrigerant
- Door heating to prevent condensation
- Adjustable fan speed
- Robust housing design with HIT sandwich technology
- Stainless-steel inner chamber
- Inner doors made of metal oxide-coated safety glass
- Two racks made of stainless steel
- Large access port with silicone plug
- Robust twin casters, with brakes at the front
- Intuitive touchscreen controller with time-segment and real-time programming
- Ergonomically adjustable control terminal
- Signal triangle with light-up status display
- Internal data logger, measured values can be read out in open format via USB
- Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
- Unit self-test for comprehensive status analysis
- Data interfaces: Ethernet, USB
- BINDER Multi Management Software APT-COM™ basic edition
Technical specifications- Temperature range: +5 °C to +60 °C
- APT.line™ preheating chamber technology
- Inverter cooling unit for energy-efficient operation
- Refrigerant: R-290 (climate-neutral)
- Door heating to avoid condensation
- Adjustable fan speed
- HIT sandwich technology housing
- Stainless-steel interior
- Metal oxide-coated safety glass inner doors
- Two stainless-steel racks
- Large access port with silicone plug
- Twin casters with front brakes
- Touchscreen controller with time-segment and real-time programming
- Ergonomically adjustable control terminal
- Signal triangle with illuminated status display
- Internal data logger; values exportable via USB in open format
- Independent temperature safety device Class 3.3 (DIN 12880) with visual and acoustic alarms
- Unit self-test function
- Interfaces: Ethernet, USB
- Included software: BINDER APT-COM™ basic edition
Downloads- Operating manual KB PRO
- Operating manual ICH light module
- Data sheet ICH light module