The Papillon+ system and its range of accessories are designed to deliver an optimal radiation dose to cancerous or suspected-cancerous tissues, while preserving healthy tissues, thanks to precise beam collimation and a precisely controlled depth of dose penetration.
This multimodality system seamlessly adapts to various clinical applications (skin, breast and rectal cancer), with parameters adjusted and tailored to indication. The use of low‑energy X‑rays reduces in facility constraints related to radiation, while maintaining high therapeutic efficacy.
Papillon+ acts as a compelling boost treatment, delivered in addition to conventional therapies.
Clinical applications
Breast cancer
Intraoperative radiotherapy (IORT)
Skin cancer
Treatment of basal cell carcinomas and squamous cell carcinomas; tumours of other histological types may also be treated, based on the clinical judgement of the lead physician.
Rectal cancer
Management of mid/low rectal adenocarcinomas, tumour size
Benefits for users
Multiple applications: breast, skin, rectal cancer
Mobile ergonomics, optimized for system re-positioning
High dose rate (up to 25 Gy/min)
Rapid treatment delivery (in less than five minutes)
Rectal indication: clinically proven efficacy through the OPERA trial
Existing French reimbursement codes for skin and rectal indications
Technical Specifications and Compliance
Maximum voltage: 50 kV
Anode current: 1.0 to 3.0 mA
Beam quality: 0.175 mm Al CDA (50 kVp)
Radiation angle: up to 310° (breast)
Weight: 130 kg
Battery‑powered operation, wireless connection between generator and computer unit
Quality management system certified to ISO 13485:2016