OverviewThe P3/P6 is a neonatal temperature control system designed for therapeutic hypothermia in neonatal hypoxic-ischemic encephalopathy (HIE). It supports rapid induction, precise maintenance and controlled rewarming according to clinical protocols to reduce mortality and limit brain injury.
Induction- Rapid cooling via water/cooling blankets to reach a target core temperature of 33 °C–34 °C within approximately 2 hours.
- Supports reproducible induction protocols suitable for neonatal intensive care units.
Cooling TechnologyThe P6 model employs Semiconductor Cooling Technology to achieve target temperatures rapidly and consistently. Cooling is delivered through water-based blankets specifically designed for neonatal application.
Control Systemi-Servo™ intelligent closed-loop temperature control: automatically adjusts water temperature based on monitored patient core temperature to maintain the prescribed target and minimise temperature deviations.
Monitoring & Safety- Simultaneous monitoring of patient core temperature, skin temperature and water temperature for continuous safety surveillance.
- Modular monitoring allows 24-hour vigilance and integration with bedside monitoring platforms.
Compatibility & Integration- Plug & play C31 compatibility with COMEN products including P6, B8 infant incubator, BQ80 radiant warmer and modular patient monitors (C series) to facilitate clinical integration.
Patient Interface & Accessories- Flat-type water blanket for targeted regional cooling.
- Full-body wrapping blanket options for uniform temperature control.
Clinical UseIndicated for therapeutic hypothermia in neonatal HIE. Supports clinical protocols for induction to 33 °C–34 °C, maintenance up to 72 hours and controlled rewarming to reduce apoptotic processes and cerebral oxygen demand.
Technical specifications- Model: P3/P6
- Indication: therapeutic hypothermia for neonatal HIE
- Induction target: core temperature 33 °C–34 °C
- Induction time: target reached within approximately 2 hours
- Treatment duration: supports protocols up to 72 hours
- Cooling method: water/cooling blankets; P6 includes Semiconductor Cooling Technology
- Control: i-Servo™ intelligent closed-loop temperature control (automatic water temperature adjustment based on core temperature)
- Monitored parameters: patient core temperature, patient skin temperature, water temperature
- Compatibility: C31 plug & play with P6, B8 incubator, BQ80 radiant warmer, C-series monitors
- Patient interfaces: flat water blanket, full-body wrapping blanket