Neonatal incubator is warming equipment used for providing constant temperature and humidity for treatment of premature infants and critically ill neonates which fit well with their physiological characteristics and needs.
COMEN B8 incubator was thoughtfully designed to provides stable temperature and suitable humidity ambience. It also features configured vital signs monitoring function to care for even the most fragile patients.
Incubator System
Intelligent Variable Resistance Heating Technology
Thanks to dual resistive heating tube and fin-styleradiator, B8 shorten heating time of the incubator to≤ 35 mins while working with its control system, which saves time in immediate care of preterms,in addtion, temperature fluctuation is significantlyreduced to 0.5 ℃.
Advanced air circulation: Symmetrical dual inlet and outlet.
A more comfortable environment with stable temperature and suitable humidity ambiance is within reach.
3 independent over-temperature protections
When the control system detects that the temperature inside the incubator exceeds limits, the heating will be stopped automatically
A sensor is applied to controltemperature through sending signals to heating instrument
A mechanical over-temperature switch is used to prevent over-temperature conditions by utilizing its high-temperature deformation characteristics.