Reducer A40 - for medical oxygen
The device is used to reduce the pressure of oxygen taken from the bottle
Destiny:
Supply of medical equipment with oxygen cylinders, pressure reduced to a value of 0.4 0.1 MPa / w. II. III /
execution I
Reducer integrated with the dispenser and the humidifier (type A21) for driving reactive oxygen therapy; is administered to a patient humidified oxygen or oxygen-air mixture; a container for sterilization (121ºC)
execution IV
Reduction to administer oxygen to a patient, sequential integrated dispenser (A41), the flow of 0-15 l / min, with an outlet to drain the patient - adjustable flow rate: 0, 2, 4, 6, 8, 10, 12, 15 l / min.
execution V
Regulator to administer oxygen to the patient primarily intended for emergency medical services, integrated dispenser sequential (A41) with a flow of 0-25 l / min, with exit to drain and quick-adjustable flow rates: 0, 2, 4, 6, 10, 15, 20 25 l / min
Technical data:
Supply pressure of 2 to 20 MPa
The pressure reduced
0.35 ± 0.5 MPa (ver. I, IV, VI)
0.4 0.1 MPa (ver. II, III, V)
Flow accuracy of ± 10% of the indicated value on the dispenser or 0.5 l / min
Gas flow rate ≥ 100l / min
Dampening gas ~ 30% RH
Operating temperature 5 ÷ 40 ° C
Connector cylinder 3/4 / "
Dimensions H x W x D:
Cast: I - 329 x 110 x 188 mm
lecture II and III - 119 x 144 x 42 mm
lecture IV 120 x 145 x 80 mm
Cast: V - 89 x 125 x 115 mm