Product overviewGEM® Premier 7000 with iQM3 is a point-of-care blood gas analyzer that integrates whole-blood hemolysis detection to identify a major source of preanalytical error and support timely clinical decisions at the bedside.
Key benefits- Detects and flags hemolysis in whole blood in approximately 45 seconds to reduce reporting of compromised samples.
- Enhances sample-quality monitoring at the point of care by addressing common preanalytical errors (hemolysis, clots, bubbles) via the iQM3 module.
- Automated GEM PAK cartridge reduces manual, labor- and skill-intensive tasks.
- Supports streamlined control and continuous quality improvement through integrated connectivity and data management options.
How whole blood hemolysis detection is possible- Acoustofluidic flow cell for rapid plasma separation and hemolysis assessment on whole blood.
- Optical detector and LED light source illuminate the optical flow cell and enable photometric absorbance calculations at each wavelength for whole-blood measurement.
GEM PAK: advanced simplicity at the POC- Automates the most labor- and skill-intensive processes to improve productivity and reduce operator variability.
- Self-contained, multi-use cartridge design limits operator biohazard and infection exposure.
- Stored at room temperature; no refrigeration required.
- Standard onboard life: 31 days (600-test PAK: 21 days); replacement schedule reduces maintenance complexity.
- No hands-on troubleshooting or corrective actions required by the operator.
- Multiple menu and test-volume configurations available to match clinical workflows.
Menu and PAK configurationsMeasured Analytes: pH, pCO2, pO2, Na+, K+, Cl-, Ca++, Glu, Lac, Hct, tBili, tHb, O2Hb, COHb, MetHb, HHb, sO2
Derived (Calculated) Parameters: BE(B), BE(ecf), tHb(c), Ca++ (7.4), Anion gap (AG), P/F ratio, pAO2, CaO2, CvO2, p50, O2cap, sO2(c), O2ect, HCO3– std, TCO2, HCO3– (actual), A-aDO2, pAO2/pAO2, RI, CcO2, a-vDO2, Qsp/Qt (est), Qsp/Qt, Hct(c), OI, Osm
Test volumes: 75, 100, 150, 300, 450, 600 µL (note: 31-day onboard use-life except 600-test PAK which has 21 days)
Menu options (examples): Blood Gas; Blood Gas, Electrolytes, Hct, tHb, O2Hb, HHb, COHb, MetHb, sO2, tBili; Blood Gas, Electrolytes, Glu, Lac, Hct, tHb, O2Hb, HHb, COHb, MetHb, sO2, tBili
Highlights- iQM3 provides rapid sample-quality assurance and hemolysis detection in ~45 seconds.
- GEM PAK cartridge consolidates complex workflows into a single self-contained unit.
- Connectivity and data-management options simplify compliance, auditing and continuous quality improvement.
Technical specifications- Product: GEM® Premier 7000 with integrated iQM3 hemolysis detection.
- Detection principle: photometric absorbance measurement on whole blood with acoustofluidic plasma separation and optical detection.
- Hemolysis detection time: approximately 45 seconds.
- Sample type: whole blood (point-of-care testing).
- GEM PAK: self-contained, multi-use test cartridge; room-temperature storage; standard onboard life 31 days (600-test PAK: 21 days).
- Measured analytes: blood gases, electrolytes, metabolites (Glu, Lac), hematocrit/hemoglobin variants and bilirubin as listed above.
- Derived parameters: extensive calculated respiratory, circulatory and acid–base parameters (see menu configurations).
- Test volumes available: 75–600 µL (multiple configurations).
- Integrated quality management: iQM3 patient-based QA integrated on the analyzer.