Endodontic micromotor control unit EndoPilot mobil
electricbenchtopcomplete set

Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set - image - 2
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set - image - 3
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set - image - 4
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set - image - 5
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set - image - 6
Endodontic micromotor control unit - EndoPilot mobil - Komet Dental - electric / benchtop / complete set - image - 7
Add to favorites
Compare this product

Characteristics

Applications
for endodontics
Operation
electric
Ergonomics
benchtop
Options and accessories
complete set

Description

The lip clip and the instrument serve as electrodes. The lip clip is hooked in the patient’s cheek pouch on the opposite side of the tooth to be treated. Make sure that the root canal is moist. Without ion flow, no measurement can take place. The canal should therefore be moistened with saline solution. The dental enamel and the cavity floor should, however, be dry in order to prevent a parallel flow and an erroneous, premature indication of the apex, i.e. at too high a position in the canal, where in fact the apex has not yet been reached. Contact with conducting filling residues can also lead to a premature indication of the apex. Has the lip clip been correctly positioned? It should be in contact with the oral mucosa. Make sure that it is not placed on the rubber dam. Does the tooth have any cracks or perforations? If so, a measurement is not possible because the apex would be indicated too early. Are there any residues of medicaments, excessive moisture or sodium hypochlorite (>2%) in the canal? If so, the presence of a significantly increased number of ions can lead to the measuring result being shifted in a coronal direction. This would lead to a false reading, i.e. the apex would be indicated too early. Prior to the measurement, extract any sodium hypochlorite solution and rinse the canal with physiological saline solution. Compare the established working length with the control radiograph. Prior to mechanical processing: • Is the root canal free from obstructions along its entire length? At the beginning of each treatment, the patency of the canal should be checked with a probe.

Catalogs

No catalogs are available for this product.

See all of Komet Dental‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.