Fitted with a zirconium dioxide sensor and a nondispersive infrared sensor, the MAT1500 is our number one in terms of its range of application, performance and precision. Measurements are completed within no more than approx. ten seconds. As with the MAT1400, highly precise results can be achieved with single measurements of oxygen concentrations thanks to the min/max detection, even if only a small sample volume is available.
Basic functions
Easy and intuitive to operate via touchscreen – even for untrained personnel
Drift-free zirconium dioxide sensor (ZrO2) measures in the measurement range range of 0-100Vol.-%
Low-drift non-dispersive infrared sensor (NDIR) measures in the measurement range range of 0-100Vol.-%
with an accuracy of ±0,5Vol.-%
Short measurement period
Only a small sample volume required
Various interfaces for the convenient transfer of measured values (USB, Ethernet, RS-232)
Any number of measurement methods for monitoring the measurement process by method, batch, product and/or production line, including limit value monitoring
Technical data
TABLE
Specification
Minimal headspace volume 2) - 5ml
Heating time - 10 minutes
Adjustment - Automatic (menu-driven)
Scales - Oxygen, carbon dioxide, nitrogen
Methods & users - Over 10000 possible
Devices basic data
Operation - 7.0” capacitive touchscreen, 800 x 480 pixels Touchscreen, keyboard, mouse, barcode scanner
Additional features - Input of product and batch designations, alarm limit management, user management,
Measured value stabilization, air pressure display
Temperature data
Ambient temperature - 10-50° C
Humididy measurement gas - <90% RH (non-condensing)