Product overviewLabotemp Semi-automatic clinical chemistry analyzer LCA-101 is a bench-top system for routine biochemical testing in hospital, clinical and research laboratories. The analyzer supports end-point, fixed-time and kinetic methods for quantitative and semi-quantitative assays on serum, plasma, urine and other biological samples. Key user interfaces include a 7.0-inch LCD touchscreen and internal storage for up to 20,000 results. An integrated incubator with selectable temperature settings provides controlled reaction conditions for repeatable measurements.
Design & operationThe LCA-101 uses a flow-through direct-reading cuvette design and a multi-wavelength optical system covering 340–630 nm for broad assay compatibility. Built-in thermal printing and RS-232/USB interfaces enable local printing and data transfer to LIS/LIMS. Its compact bench-top footprint, low power consumption and vibration-resistant construction make it suitable for small clinical labs and higher-throughput diagnostic environments.
Why choose LCA-101- Multi-method testing including end-point, fixed-time and kinetic protocols
- Stable temperature control for consistent reaction conditions
- Intuitive touchscreen operation with large on-board data storage
Features- 7.0-inch LCD touchscreen for user-friendly control
- Advanced optical system with stable light source for consistent absorbance across wavelengths
- Multi-mode calibration (linear and non-linear curve fitting)
- Real-time reaction curve display to monitor assays
- User-defined test parameters for custom assays and reagents
- Automatic blank correction to reduce manual errors
- Internal QC program supporting daily validation and compliance workflows
- Low reagent consumption design to optimize operating costs
- Compact bench-top structure with vibration resistance
- Quick warm-up system to reduce waiting time
Applications- Routine and semi-quantitative biochemical assays (enzymes, metabolites, glucose, lipids)
- Temperature-controlled, multi-wavelength photometric testing for hospital diagnostics, clinical laboratories, pharmaceutical QC and research facilities
Technical specifications- Incubator temperature: R.T., 25°C, 30°C, 37°C (four options)
- Cuvette: Flow-through direct-reading cuvette
- Wavelengths: 340 / 405 / 450 / 510 / 546 / 578 / 630 nm, plus 2 open filter positions
- Wavelength accuracy: ±2 nm
- Screen: 7.0 inch LCD touch screen
- Language: English, French, Spanish (customizable)
- Printer: Built-in thermal printer
- Interface: RS-232, USB
- Power consumption: 150 VA
- Power supply: AC 110/220V ±10%, 50/60 Hz
- External size (W×D×H): 400 × 410 × 190 mm
- Packing size (W×D×H): 520 × 490 × 330 mm
- Net weight: 9 kg
- Gross weight: 12 kg
- Incubating positions: 20 incubating positions
- Sample type: Serum, plasma, urine, cerebrospinal fluid, etc.
- Analytical methods: End point, Fixed time (two-point), Dynamic (rate) method