1. Products
  2. Turbidity meter
  3. Labotemp Systems

Laboratory turbidity meter LTM-02
portable

Laboratory turbidity meter - LTM-02 - Labotemp Systems - portable
Laboratory turbidity meter - LTM-02 - Labotemp Systems - portable
Add to favorites
Compare this product

Characteristics

Options
laboratory, portable

Description

Product Overview
Labotemp Portable Turbidity Meter LTM-02 is a compact turbidity meter for accurate on-site and laboratory analysis of water clarity. It is suitable for drinking water verification, pool monitoring, environmental field surveys and industrial liquid process checks. The instrument uses an infrared LED (850 nm) and a silicon photodiode detector to provide stable, low-interference measurements across a wide range.

Operation & Functionality
The meter supports 2–5 point calibration with formazin standards and selectable units (NTU, FNU, EBC, ASBC). Measurement modes include single and continuous readings, automatic stable-reading lock and internal memory for up to 100 records. Data export is available via USB. Power options: 3 × 1.5V AA batteries or DC 5V adapter for extended field use.

Key Benefits
  • Compact handheld design for field and laboratory use
  • Flexible calibration: 2–5 points with multiple turbidity units
  • Stable readings using infrared LED and silicon photodiode detector
  • Data management: 100-record memory and USB communication
  • Single and continuous measurement modes with automatic stable lock
  • Dual power: battery operation or DC 5V adapter

Applications
  • Municipal water distribution turbidity checks
  • Swimming pool and recreational water monitoring
  • Environmental on-site assessments and field surveys
  • Industrial liquid process monitoring and quality control
  • Academic and research portable sample analysis

Technical Specifications
  • Measuring Range: 0–1100 NTU; 0–275 EBC; 0–9999 ASBC
  • Resolution: 0.01 NTU (0–99.9 NTU); 0.1 NTU (100–999 NTU); 1 NTU (1000–1100 NTU)
  • Accuracy: ±5% of reading (0–500 NTU); ±8% of reading (501–1100 NTU)
  • Calibration: 2–5 points
  • Calibration Standards: 0.02, 10, 200, 500, 1000 NTU
  • Light Source: Infrared LED (850 nm)
  • Detector: Silicon photodiode
  • Stray Light: <0.02 NTU
  • Sample Volume: 30 ml
  • Memory Capacity: Stores up to 100 data sets
  • Output Interface: USB communication interface
  • Display: Custom LCD (60 × 40 mm)
  • Power Supply: 3 × 1.5V AA batteries or DC 5V adapter
  • Auto Power Off: 15 minutes after last key press
  • Dimensions (L×W×H): 180 × 85 × 70 mm
  • Package Size: 30 × 10 × 27 cm
  • Net Weight: 0.3 kg
  • Gross Weight: 1.5 kg