Product OverviewDisposable hemoperfusion cartridge designed for blood purification and support of hemodiafiltration in patients with acute or chronic renal failure; adsorption-based removal of target toxins and inflammatory mediators.
Product Features- Housing: polypropylene (PP), BPA-free
- Adsorbent: polystyrene divinylbenzene macroporous resin with resin coating technology for improved biocompatibility
- Optimized pore diameter for removal of medium and large molecules
- High mechanical strength and low particle shedding
How the Disposable Hemoperfusion Cartridge WorksBlood flows through the cartridge where the internal adsorption media binds target toxins and inflammatory mediators by adsorption; suitable for integration in hemoperfusion or combined hemodiafiltration circuits (illustration on original page).
Ordering Information (Models & Adsorbent Volumes)Model: OC-60(A), OC-80(A), OC-90(A), OC-100(A), OC-110(A), OC-120(A), OC-130(A), OC-140(A), OC-150(A)
Adsorbent volume (mL): 60, 80, 90, 100, 110, 120, 130, 140, 150
Model: OC-160(A), OC-170(A), OC-180(A), OC-190(A), OC-200(A), OC-230(A), OC-240(A), OC-250(A), OC-280(A)
Adsorbent volume (mL): 160, 170, 180, 190, 200, 230, 240, 250, 280
Model: OC-300(A), OC-330(A), OC-350(A), OC-380(A), OC-400(A), OC-430(A), OC-450(A), OC-480(A)
Adsorbent volume (mL): 300, 330, 350, 380, 400, 430, 450, 480
Components and MaterialsAdsorbent: polystyrene divinylbenzene macroporous adsorption resin.
Cylinder, end cover, cap, net rack: polypropylene (PP).
Filter mesh: Nylon-66.
Filling liquid: injection water.
Sealing plug, sealing ring: silicone rubber (SIR).
Adsorption PerformancePentobarbital sodium ≥ 80%; Creatinine ≥ 15%; VB12 ≥ 90%; β2-MG ≥ 80 μg/mL; PTH ≥ 7.0 ng/mL; IL-6 ≥ 600 pg/mL.
Operational ParametersBlood flow rate: 100–250 mL/min.
Pressure resistance: 100 kPa.
Sterilization method: irradiation sterilization; shelf life: 3 years.
Test ConditionAdsorption tests for pentobarbital sodium, creatinine, VB12: (60 ± 10 cycles/min, 37 ± 1°C). Adsorption tests for β2-MG, PTH, IL-6: (140 ± 10 cycles/min, 37 ± 1°C).
Technical specifications- Product type: disposable hemoperfusion cartridge for blood purification/hemodiafiltration support
- Model family: OC series (multiple OC-xxx(A) variants with different adsorbent volumes)
- Adsorbent material: polystyrene divinylbenzene macroporous resin
- Housing & structural materials: polypropylene components; Nylon-66 filter mesh
- Sealing materials: silicone rubber (SIR)
- Filling fluid: injection water
- Adsorption efficacy examples: pentobarbital sodium ≥ 80%, VB12 ≥ 90%, β2-MG ≥ 80 μg/mL
- Recommended blood flow rate: 100–250 mL/min
- Pressure resistance: 100 kPa
- Sterilization: irradiation sterilization; shelf life: 3 years
- Test temperatures and flow conditions specified for performance evaluation