mobile laboratory incubator / heating / cooling / stainless steel
ICP110

Add to MyMedicalExpo favorites
Compare this product Don’t compare this product
mobile laboratory incubator / heating / cooling / stainless steel mobile laboratory incubator / heating / cooling / stainless steel - ICP110
{{requestButtons}}

Characteristics

  • Configuration:

    mobile

  • Other characteristics:

    heating, cooling, stainless steel

  • Temperature range:

    Max.: 60 °C (140 °F)

    Min.: -12 °C (10.4 °F)

  • Capacity:

    108 l (28.5 gal)

Description

Compressor-cooled incubator ICP110 Ideal for ramp operation with rapid temperature changes during incubation, breeding or storage.

Compressor-cooled incubator ICP110
Standard delivery Compressor-cooled incubator
With the help of our model selection, with dimensioned model sketches and extensive technical data for download, you can find the right cooled incubator ICP for your needs. For small volumes and for work predominantly in the proximity of the ambient temperature, the Memmert Peltier-cooled incubator is recommended. Flexibility and technical features of our appliances meet all possible needs. Put us to the test!

Sketch Compressor-cooled incubator ICP110
Temperature
Setting temperature range-12 to +60 °C
Working-temperature rangefrom -12°C up to +60°C (Optimum performance of cooling aggregate at an ambient temperature of +16°C up to +34°C. Not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over.)
Setting accuracy temperature0.1 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Control technology
ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable temperature values
adjustable parameters