What is 123 KIT
123 KIT features
123KIT reduces the number of conventional drilling sequence to once in soft bone, to twice in normal
bone, and 3 times in hard bone. Both patients and physicians can both have a win-win by shortening
the time of implant surgery. The 123Kit can also maximize the convenience for the users by integrating
drilling sequences for TS, SS, and US Fixture.
1. Speedy Drilling Sequence: Only three times
(TSII 5.0 Fixture: Hard bone)
2. Simple & Easy Sequence: Unification of drilling sequence
(Separate drilling sequence by system Unified drilling sequence by type.)
3. Design of drill stop
: Drill Stop designed with fixture lengths, the drilling depth can easily be adjusted safely and conveniently.
① During surgical procedure, keep used tools in saline solution or
distilled water.
② When the surgery has been completed, soak the used tools in alcohol
for cleaning.
Washing with hydrogen peroxide is prohibited.
Exposure to hydrogen peroxide may discolor the laser
marking and TiN coating.
③ Wash blood stains and other foreign matter clean with distilled water
or flowing water.
④ Remove moisture with a dry cloth or a hot air blower.
⑤ Place the dried tools in the KIT case.
(Refer to the color coding for placement of tools in the kit case.)
⑥ After placement, sterilize the kit in an autoclave at 132℃ for 15
minutes and store at room temperature