External card to expand the memory to 10,000 cases
Automatic analysis and automatic evaluation
Internal rechargeable lithium ion battery
Paper speed: Automatic & manual recording: 6.25, 12.5, 25, 50mm/s, Error: ±5%
USB interface (standard), LAN socket (option)
Built in memory with large capacity, can store more than 1000 cases
Optional PC software for data transfer (save, review, print a report)
Re analysis and display of 12 leads
Analysis elements for FCG, HFCG, QTD and HRV
Multiple languages
The device is supplied with a carrying case
Technical Specification:
Lead: Standard 12 leads;
Input circuit: protection against defibrillator;
Acquisition Mode: Simultaneous 12 leads acquisition and asynchronous acquisition;
Measurement Range: ±5mVpp;
Time Constant: ≥3.2s;
Frequency Response: 0.05Hz~150Hz (-3dB);
Sensitivity: 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV;
Input Impedance: ≥50MΩ;
Input Circuit Current <0.05µA;
Noise Level <15µVp-p;
Patient Current Leakage <10µA (220V-240V);
Filter:
EMG Filter: 25 Hz / 30 Hz / 40Hz
Low Pass Filter: 75 Hz / 100 Hz / 150Hz
DFT Filter: 0.05 Hz/0.15 Hz/0.25 Hz/0.32 Hz/ 0.5 Hz/ 0.67Hz/ ADS
AC Filter: 50 Hz / 60Hz
CMRR: 100dB;
Recording Mode:
Automatic: 3CH×4+1R, 3CH×4,
Manual: 4CH, 3CH, 2CH, 1CH
Rhythm: Any lead selectable
Recording System: Thermal matrix printing;
Recording Paper: Roll 80mmx20M;
Paper Speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s;
Interface: Network interface, USB Master/Slave, SD card;
Safety Standard: IEC I/CF;
Sample Rate Normal: 1000sps/channel
Power Supply:
AC: 100~240V, 50/60Hz, 30VA~100VA
DC: 14.8V/2200mAh, built-in lithium battery