The REM 4000 captures high-quality photos and automatically analyses the size, shape and population of the endothelial cells. The monitoring of contact lens wearers and evaluation of common ocular conditions becomes an easy routine.
Product Features
13 measurement areas
Counts more than 300 cells
Integrated non-contact Pachymetry
Automatic analysis, L-count, Core method, Dark Area analysis
Integrated database
Auto alignment + auto measurement
Specification
RESOLUTION
Pixels used for picture taking
480 (V) x 180 (H) pixels
Capturing scope
0.25 x 0.54 mm
1 centre + 12 peripheral measurements
13 x fixation points
Min. cell resolution
1.14 μm (V) x 1.45 μm (H)
Optical magnification
x 228
Display
10.4’’ LCD Colour
Display resolution
1.14 μm
MEASUREMENT
Auto alignment
Yes
Auto measurement
Yes
Manual mode (1 & 2)
Yes
MEASUREMENT FUNCTION
Automated captured examination
16 pictures for analysis
Number of analysed cells
More than 300 cells
Capturing position
Center + 12 peripheral points
Analysis method
Automatic analysis, L-count, Core method, Dark Area analysis
Analysis values
CD (cell density), AVG (average cell area), SD (standard deviation of cell area)
CV (coefficient of variation of cell area), Cell size (max. + min. cell area)
Stroke of moving section
X: 88 mm, Y: 40 mm, Z: 50 mm
Stroke of electrical chin rest
70 mm
Measuring accuracy Pachymetry
+/- 10 μm
OPERATING ENVIRONMENT
Temperature
+10° to +40°
Humidity
30 % to 75 %
Atmospheric pressure
800 to 1060 hPa
Standards applied
MDD Annex ii, iSo 13485
DATA MANAGEMENT
Built-in printer
Thermal printer
Data output type
USB-Hx2, USB-Dx1, LAN, SD Card (for internal database)