Constant current, multimode stimulator for direct cortical electrical stimulation.
SGM-STIM-2 stimulator is designed to be used for direct electrical stimulation of cerebral cortex focal regions for motor, sensory and language functions. Constant current types of stimulus are used to deliver controllable amount of electrical charge, regardless of the impedance of tissue between electrodes.
Stimulator can be used as a stand-alone device with full local functionality, or can be connected and remotely controlled from IOM system or remote computer.
Major Features
High intensity constant current multipulse stimulator.
Current range: 0 to 40 mA.
Selectable stimulating train rates 0,1 to 8,0 Hz in steps of 0,1 Hz (externally and internally controlled, “trigger in” and trigger out”).
Number of pulses in the train: 1 to 9.
Interstimulus time: 1-10 min 0,1 ms steps.
Power supply voltage 110/220 V automatically detected.
Set of necessary accessories including Output Extender SGM S2-840.
Stimulation probes used: eightpole strip electrodes or handheld probe connection interface.
pecifications
1.3.1 STIMULUS OUTPUT -
Current range - 0 - 5 mA in 0.2 mA steps / 5 – 40 mA in 1 mA steps
Pulse width - 100 - 1000 microseconds square wave (100 µs step)
Maximum output voltage - 50 V
1.3.2 STIMULUS PULSE RATES -
Pulse or group rate - 0.1 to 8.0 per second, in 0.1 steps
Number of electrodes - 9
Burst selections - Number of pulses – 1 to 9
Interstimulus time - 1 - 10 ms in 0.1 ms steps
1.3.3 TRIGGER FACILITIES -
Manual - Push button mounted on the front panel
Auto -
External - TTL compatible positive edge logic signal
Trigger output - Positive TTL compatible signal 500 µs wide