Analysis method Two-point, end-point, kinetic, spline
Analysis item 200 colorimetric items
Sample & Reagent system
Sample & reagent probe
1 independent probe with stirring function
Sample cup specification
Standard cup Ø12X37mm
Blood tube Ø12X100mm
Plastic tube Ø12X100mm
Sample & reagent probe technology
Liquid level detection,
X+Y+Z dispensing system
Sample & reagent probe cleaning
Automatic washing both interior and exterior
Reagent bottle
15ml
Reaction system
Reaction position
48, optical cuvettes
Cuvettes specification
Optical diameter is 5.8mm
Reaction temperature
37℃
Wastewater treatment
Waste liquid level alarming (optional)
Purified water level alarming (optional)
Reaction cuvette cleaning
4-step auto water washing with detergent
Optical system
Light source
10W/6V halogen lamps
Spectrophoto metry
Forward
Wavelength
8wavelengths: 340nm, 405nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 700nm
Absorption range
0~3.5Abs
Resolution
0.0001Abs
Calibration and QC
Calibration method
1-Point linear, 2- Point linear, multiple point linear, non-linear method
QC method
Real-time QC days QC & day QC
Control rule
Westgard multi-rule, L-J plot
Operating system
Operating system
Windows XP /Windows 7