Overview
Spellman’s Injector Control Assembly (ICA) multi-output high voltage power supplies provide the electrode voltages for a triode electron gun to inject electrons into a Linear Accelerator (LINAC). The resulting high-energy electron pulses generate high-energy X-Rays used for applications such as cargo screening and cancer therapy. This rack-mount multi-output supply provides Cathode, Heater and Grid voltages; the Cathode current can be pulsed up to 500 Hz. Heater and Grid supplies float at the Cathode voltage potential, typically up to -15 kV. Remote control and monitoring are available via standard Ethernet protocols. Front-panel LED indicators provide basic output and fault status, including Cathode Current and Arc faults.

Key Features
  • Multi-output electron-gun power supply
  • Fast rise-time pulsed cathode current (up to 500 Hz)
  • Remote programming and monitoring via Ethernet
  • Customizable for OEM and system integration


Typical Applications
  • Cargo screening
  • Cancer therapy (LINAC)


Indicator Panel
INDICATOR | COLOR | CONDITION (illuminated when...)
Trigger Present | Green | Input trigger is present.
AC Power | Green | AC supply voltage is present.
Interlock In | Green | Interlock in signal is present.
Arc Fault | Red | Arc fault signal is active.
Cathode Current Fault | Red | Cathode current fault signal is active.
Cathode Voltage On | Amber | Cathode high voltage is present.

Datasheet Note
*All specifications are subject to change without notice. Refer to the product datasheet for the most up-to-date revision.

Technical specifications
  • Reference: Ref. 128129-001 REV. C
  • Input Voltage: 100-240 Vac, 50-60 Hz
  • Cathode Output Voltage: 0 to -15 kV
  • Pulsed Cathode Current: 200 mA to 1500 mA
  • Mean Cathode Current: 5.5 mA (max)
  • Current Pulse Width: Up to 6 μs (90% to 90%)
  • Current Pulse Frequency: Up to 500 Hz
  • Current Risetime: 200 ns (typical)
  • Heater Output Voltage (wrt Cathode): 0 to -6.5 Vdc
  • Heater Current: 3 A
  • Heater Resistance: 0.7 ohm
  • Grid (fixed) Voltage (wrt Cathode): -60 to -135 Vdc
  • Grid (fixed) Current: 250 mA
  • Pulse Voltage: 0 to 340 Vdc (wrt Cathode) — Injector current is regulated by programming the Grid Pulse Voltage
  • Connections — Front Panel: Ethernet (RJ45 jack); Interlock (24 V nominal, 9 pin D-sub male pin plug); Pulse Input (BNC)
  • Connections — Rear Panel: AC Power In (IEC320 C14 male pin receptacle); HV Out (3 pin terminal block)
  • Interlock: 24 V must be present to enable output high voltage cathode; intended to connect to system interlocks to disable the ICA if system interlocks are not satisfied
  • Operating Temperature: 10°C to +40°C
  • Storage Temperature: -40°C to +70°C
  • Humidity: 30% to 60% relative humidity, non-condensing
  • Weight: 15 lbs (6.8 kg)
  • Regulatory Approvals: Designed in accordance with IEC 60601-1 (Edition 3) requirements; UL 94V-0 and RoHS compliant