Bidirectional
Helps protect against oxygen tubing fires
A thermal fuse designed to stop the flow of gas in the event that the downstream oxygen tubing is ignited
Helps keep patients safe no matter what method of oxygen delivery they use
Offers a cost-effective solution to lower the risk of serious patient injury in an oxygen fire
Maximum Operating Pressure 600 kPa*
*Maximum static pressure delivered by the gas supply source
Flow Rate0.25* 25 l/min
The Firesafe Cannula Valve can be safely used at lower flow rates but may not meet the internal leak rate of 10 ml/min when activated in the event of a fire.
Resistance to Flow ≤ 0.07 kPa at 2 l/min
≤ 0.3 kPa at 5 l/min
≤ 1 kPa at 10 l/min
Intended Life 5 years, expiry date engraved on device
Application Single patient multiple use
Packaging - 100 units per box (5 bags per box, 20 units per bag)
Connections EN 13544-2, 6 mm nominal tubing connection
Environmental Limits for Transport and Storage Temperature: -25 °C to 70 °C (-4 °F to 140 °F) Humidity: 0 – 100 % RH
Environmental Limits for Use Temperature: 0 °C to 50 °C (32 °F to 122 °F) Humidity: 0 – 100 % RH
Applied Standards EN 13544-2:2002 – Respiratory Therapy Equipment – Part 2: Tubing and connectors.
Regulatory - EC: MDD Class IIa Medical Device
USA: FDA Registered Class 1 Medical Device; 510(k) exempt
GMDN Code and Term: 60391 – Fire safety valve