Microbial enumeration system MD500 offline

Microbial enumeration system - MD500 offline - Tailin
Microbial enumeration system - MD500 offline - Tailin
Microbial enumeration system - MD500 offline - Tailin - image - 2
Microbial enumeration system - MD500 offline - Tailin - image - 3
Add to favorites
Compare this product

Description

Introduction
The MD500 Offline Microbial Detection Analyzer combines laser-induced fluorescence and Mie scattering particle measurement to count individual microorganisms in relatively pure water streams. Designed for continuous bioburden monitoring, it delivers real-time alerts without waiting for culture plate results and requires no reagents or sample preparation.

Key features
  • Measurement range 0–10,000 AFU/mL with a detection limit of 1 AFU and minimum particle detection size of 0.1 μm.
  • Reagent-free operation with full touchscreen interface for straightforward on-site use.
  • Multiple output formats: configurable 4–20 mA, analog/digital channels, Ethernet/Modbus TCP, USB and Wi‑Fi for SCADA integration.
  • Built-in methods and automated maintenance: one-key flushing, continuous monitoring of up to 10 channels (expandable to 100).
  • Designed for pharmaceutical and process environments: supports GMP and 21 CFR Part 11 requirements with four levels of password permissions, audit trail and multiple backup options.


Applications
  • Inline monitoring of injection water and purified water.
  • Rinse water and product solution control.
  • Ultra-pure water and natural drinking water surveillance where low-burden detection is required.


Technical specifications
  • Measurement range: 0–10,000 AFU/mL
  • Minimum detection size: 0.1 μm
  • Detection limit: 1 AFU
  • Response time: real-time
  • Analysis mode: continuous operation
  • Sample temperature: 5–90 ℃
  • Flow rate: 1.5–30 mL/min
  • Ambient conditions: 0–50 ℃, up to 80% RH
  • Channels: 1–10 supplied, expandable to 100
  • Power supply: 100–240 VAC, 50/60 Hz
  • Inlet connection: 1/8 inch (3.2 mm) OD
  • Weight: 10 kg
  • Data outputs: configurable 4–20 mA, analog/digital outputs, Ethernet/Wi‑Fi, Modbus TCP, USB


Compliance & data management
  • Audit-ready design to support GMP and 21 CFR Part 11 compliance.
  • Four user permission levels and full audit trail for secure data integrity.
  • Multiple backup and export options; SCADA connectivity via Modbus TCP for centralized monitoring.


Options & installation
  • Expandable channel architecture to scale from single-channel monitoring to up to 100 channels.
  • Multiple signal interfaces for easy integration with plant control systems.
  • Compact footprint and reagent-free design simplify installation and reduce routine consumables.

Catalogs

No catalogs are available for this product.

See all of Tailin‘s catalogs

Exhibitions

Meet this supplier at the following exhibition(s):

CPHI China
CPHI China

16-18 Jun 2026 Shanghai (China) Hall N3 - Stand B56

  • More information
    FILTECH
    FILTECH

    30 Jun - 02 Jul 2026 Cologne (Germany) Hall 7 - Stand N21

  • More information

    Other Tailin products

    Microbial Monitoring System

    *Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.