Determination principleReflectance and color analysis; calculates concentration values of related items using fully digital acquisition and calculation; outputs semi-quantitative results.
Test items- Glucose (GLU)
- Bilirubin (BIL)
- Specific gravity (SG)
- pH
- Ketone (KET)
- Occult blood (BLD)
- Protein (PRO)
- Urobilinogen (URO)
- Nitrite (NIT)
- Leukocyte (LEU)
- Ascorbic acid (VC)
- Creatinine (CRE)
- Urine calcium (CA)
- Microalbumin (MCA)
- A:C
Notes on test itemsThe specific items tested are determined by the matching urine test strip and can be freely combined; the A:C format can be freely selected.
Performance and features- Test speed: up to 120 tests/hour (fast continuous); 60 tests/hour (single step, slow continuous)
- Light source: LED cold light source; life ≥ 1500 hours
- Report mode: LCD display; serial output; built-in 57 mm thermal printer; Chinese and English optional
- Instrument configuration: supports 5 or more languages to meet needs of customers in different countries
- Display: LCD TFT true color screen
- Input method: full-screen capacitive touch; can be connected to a barcode scanner
- Report storage: large capacity; can store 10,000 test reports
- Power-off protection: automatically saves all measurement data on power loss
- External interface: standard RS-232 serial port or Bluetooth transmission for computer/LIS integration
- Test strip handling: test strip can be placed arbitrarily without stopping the test process; device gives an alarm in case of abnormality
- Repeatability and stability: coefficient of variation (CV) < 1.0%
- Accuracy: the difference with the labeled value of the corresponding reference solution is not more than one order of magnitude
- Automatic corrections: automatically corrects influence of urine pH on specific gravity test results; built-in temperature detection and result correction according to temperature
- Working voltage: AC 100-240 V, 50 Hz/60 Hz (wide power supply)
- Fuse: φ5 mm × 20 mm, T2AL250V
- Instrument power consumption: ≤ 35 VA
- Safety classification: basic insulation, transient overvoltage class I, pollution degree 2
- Construction: full mold structure; stable, reliable, anti-corrosion and durable
- Appearance: white, simple and generous design
Technical specifications- Model: YE-100P
- Determination principle: reflectance and color analysis (digital acquisition and calculation)
- Test items: GLU, BIL, SG, pH, KET, BLD, PRO, URO, NIT, LEU, VC, CRE, CA, MCA, A:C (dependent on test strip)
- Test speed: 120 tests/hour (fast continuous) / 60 tests/hour (single step)
- Light source: LED cold light, life ≥ 1500 h
- Report: LCD display, serial output, built-in 57 mm thermal printer; language options
- Display: TFT true color LCD
- Input: full-screen capacitive touch; barcode scanner support
- Storage: up to 10,000 test reports
- Power-off protection: automatic data save on power loss
- Interfaces: RS-232 and Bluetooth
- Repeatability (CV): < 1.0%
- Working voltage: AC100-240V, 50/60Hz
- Fuse: φ5 mm × 20 mm, T2AL250V
- Power consumption: ≤ 35 VA
- Safety: basic insulation; transient overvoltage class I; pollution level 2
- Dimensions: 200 mm × 137 mm × 52 mm
- Net weight: 600 grams
- Construction: full mold, anti-corrosion, durable
- Color/appearance: white