Class II microbiological safety cabinet / laboratory / floor-standing / vertical laminar flow
BMB-II-Laminar-S NEOTERIC LAMSYSTEMS GmbH

Characteristics

  • Class:

    class II

  • Applications:

    laboratory

  • Configuration:

    floor-standing

  • Other characteristics:

    vertical laminar flow, with HEPA filter

Description

Microbiological Safety Cabinet Class II (Type A2)
BMB-II-"Laminar-S." NEOTERIC

Available in sizes: 900, 1200, 1500 and 1800 mm.

The cabinet is designed for physical isolation (containment and controlled removal from the working zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the working room and to protect the environment, along with protection of the working agents inside the working zone from external and cross-contamination.

The cabinet does not protect against toxic chemical substances and radionuclide, and does not retain odors of the working agents.
If this cabinet is used for work with pathogenic biological agents (PBA) and microorganisms belonging to the highest (most hazardous) groups of pathogenicity, it is necessary to use an appropriate protecting (antiplaque) suit.

The cabinet is used to equip individual working places in virologic and bacteriological laboratories of medical, pharmaceutical and other institutions working with pathogenic agents.

Vertical downward unidirectional (laminar) airflow.

Airflow in the work opening directed into the front table grille, creating an air curtain