OverviewF26 SPEED is a compact leak tester designed for high-speed production and for integration in automatic or semi-automatic test machines. It performs pressure-decay leak testing up to 10 bar with selectable ΔP measurement ranges from 500 Pa to 50 kPa and offers multiple communication and networking options.
Applications- Blister packaging
- Sachets and pouches
- Pneumatic and hydraulic fittings
- Valves, connectors and molded parts
Features- Leak testing by pressure drop (up to 10 bar)
- ΔP measurement ranges: 500 Pa to 50 kPa
- 8 test programs
- Networking of several modules via CAN
- On-line setup via USB port
- Compact enclosure (H 163 x W 88 x D 85 mm)
- Flexible mounting styles
- Integrated fill and dump valves
- Flexible communications: RS232, USB, Ethernet (for printer and PC)
- Environment: RoHS compliant
Measurement ranges (table)- Vacuum: pressure accuracy ±2% of full scale; max resolution 1 mbar. ΔP measurement ranges: 500 Pa – 1 kPa – 2 kPa – 5 kPa; ΔP accuracy ±4% FS; ΔP resolution 1 Pa – 10 Pa.
- -500/+500 mbar: pressure accuracy ±2% FS; resolution 1 mbar. ΔP ranges: 500 Pa – 1 kPa – 2 kPa – 5 kPa; ΔP accuracy ±4% FS; ΔP resolution 1 Pa – 10 Pa.
- 0–500 mbar: pressure accuracy ±2% FS; resolution 1 mbar. ΔP ranges: 500 Pa – 1 kPa – 2 kPa; ΔP accuracy ±4% FS; ΔP resolution 1 Pa.
- 0–2 bar: pressure accuracy ±2% FS; resolution 10 mbar. ΔP ranges: 1 kPa – 2 kPa – 5 kPa – 10 kPa; ΔP accuracy ±4% FS; ΔP resolution 1 Pa – 10 Pa.
- 0–5 bar: pressure accuracy ±2% FS; resolution 10 mbar. ΔP ranges: 2 kPa – 5 kPa – 10 kPa – 20 kPa; ΔP accuracy ±4% FS; ΔP resolution 1 Pa – 10 Pa.
- 0–10 bar: pressure accuracy ±2% FS; resolution 10 mbar. ΔP ranges: 5 kPa – 10 kPa – 20 kPa – 50 kPa; ΔP accuracy ±4% FS; ΔP resolution 10 Pa.
- Optional external mechanical pressure regulator: Vacuum – 500 mbar – 5 bar – 9 bar.
TechnologyThe F26 SPEED uses a standard pressure-decay method: it compares the pressure inside the tested part with atmospheric pressure. A piezo sensor measures small pressure variations (pressure drop). This method is appropriate when the application does not require ultra-high accuracy or extremely short cycle times. The test pressure is applied to the part inlet; measurement is performed by the pressure sensor. A desensitized mode is available for specific configurations.
FactsheetsDetailed information about the F26 SPEED, including full features and specifications, is available in the product factsheet (English).
Technical specifications- Weight: around 1.5 kg
- Compact enclosure dimensions: H 163 x W 88 x D 85 mm
- External electrical supply: 24 V DC / 1.6 A; main adapter included 110–230 V AC
- Operating temperature: +10°C to +45°C
- Storage temperature: 0°C to +60°C
- Indicator lights: cycle in progress, CAN, Ethernet, 24 V
- Air supply: clean and dry air; comply with ISO 8573-1
- Communication: RS232, USB, Ethernet (for printer and PC)
- Programs: 8 test programs; networking of modules via CAN; USB setup
- Measurement capability: pressure-drop test up to 10 bar with multiple selectable ranges and resolutions
- Standards: RoHS compliance