OverviewThe F28 LPV, a vacuum decay instrument from ATEQ's F28 range, is a compact pressure-decay leak tester designed to be mounted close to the test part to ensure optimal measurement conditions. The instrument is configured and supervised via the F28 Control interface.
Key features- Pressure decay leak measurement: vacuum mode
- Water leak detection by evaporation under strong vacuum
- Smart pressure regulation without external regulator – auto fill (optional)
- Electronic regulation (optional)
- Measurement units: Pa, Pa/s, sccm, μg/min of H2O
- Compact footprint: H 141.5 x L 225 x D 97 mm
- Integrated fill and dump valves
- Ethernet network up to 50 modules (Modbus IP)
- DIN-rail mounting for compact industrial integration
- Environment: ROHS compliant
Measurement characteristicsPressure drop measurementTest Pressure (Absolute) | Range: 0.05 to 110 kPa | Resolution: 1 Pa | Accuracy*: 2% F.S.
Vacuum Decay | Range: 2000 Pa | Resolution: 1 Pa | Accuracy*: 2% F.S.
*Linearity + repeatability + hysteresis
Applications- Packaging
- Blister inspection
- Watches and micro-mechanics
- Mobile phones and small electronics
- Pneumatic and hydraulic fittings
TechnologyStandard pressure decay compares the pressure in the part with atmospheric pressure and measures the pressure variation using a piezo sensor. This method is suitable for applications that do not require extremely high accuracy or very short cycle times. The test pressure is applied to the part and the pressure sensor performs the measurement.
Technical specifications- Weight: around 3.5 kg
- Overall dimensions (mm): H 141.5 x L 225 x D 97
- Measurement units: Pa, Pa/s, sccm, μg/min of H2O
- Pressure ranges: Test Pressure (Absolute) 0.05 to 110 kPa; Vacuum Decay 2000 Pa
- Resolution: 1 Pa
- Accuracy: 2% F.S. (linearity + repeatability + hysteresis)
- Electrical supply: 24 V DC / 1.6 A; mains adapter included 110–230 V AC
- Air supply: clean, dry air required; apply air quality standard (ISO 8573-1); test pressure input may require an external regulator
- Operating temperature: +5°C to +45°C; Storage: 0°C to +60°C
- Integrated fill and dump valves
- Smart pressure regulation without regulator – auto fill (optional); electronic regulation optional
- Ethernet network support up to 50 modules (Modbus IP)
- DIN-rail mounting
- Environment: ROHS standard