SubtitleDesigned for low- to mid-volume clinical laboratories requiring automated random-access chemistry testing.
OverviewThe DxC 500 AU is a fully automated clinical chemistry analyzer providing random-access testing with STAT capability. The system standardizes user experience, test menus and quality control to support consistent results across a laboratory network. It supports ready-to-use reagents, continuous sample loading and a refrigerated reagent carousel for consolidated reagent storage.
Key benefits- Standardized user interface, menus and protocols to reduce operator variability and simplify training
- Barcoded reagents and consolidated reagent storage to facilitate inventory control and traceability
- Continuous loading and integrated STAT handling to maintain workflow during variable sample inflow
Workflow and sample handlingDxLAB sample racks accept primary tubes and nested cups for calibrators, QC and patient samples in a single rack. Continuous loading allows processing of samples as they arrive without pre-sorting. A refrigerated STAT carousel accommodates urgent samples alongside routine runs.
Reagent managementRefrigerated reagent carousel with 76 positions supports 15 mL, 30 mL and 60 mL bottles. Barcoded reagents provide automatic tracking of available tests, expiration dates and lot numbers. Reagent volume range: 10 µL–250 µL (1 µL increments).
Performance highlights- Throughput: up to 400 photometric tests/hour; up to 800 photometric tests/hour with ISE configuration
- ISE maximum: up to 600 tests/hour if configured for ISE only
- HbA1c maximum: up to 65 tests/hour if configured for HbA1c only
Main specificationsAnalytical method: Fully automated random-access clinical chemistry with STAT capability
Analytical principles: Spectrophotometry, potentiometry; colorimetry, turbidimetry, latex agglutination, indirect ISE
Assay types: Endpoint, rate, fixed point and indirect ISE
Chemistry menu capacity: >135 preprogrammed photometrics (including LIH + HbA1c and ISE)
Throughput details: 400 photometric tests/hour (up to 800 with ISE); ISE max 600 tests/hour; HbA1c max 65 tests/hour
Sample types: Serum, plasma, urine, CSF, whole blood (assay dependent)
Sample capacity/management: 168 sample tubes (24 racks × 7) with continuous loading; refrigerated STAT carousel: 22 samples concurrently
Sample container sizes: Primary/secondary tubes Ø 11.5–16 mm; height 55–102 mm; nested micro cups and sample cups
Sample volume: 1.0 µL–25 µL (0.1 µL increments)
Barcode symbology: NW7, EAN13, CODE 39, CODE 128, ISBT-128, 2 of 5 standard, 2 of 5 interleaved mixed
Reaction: Total volume 90–350 µL; reaction time up to 8 min 37.5 s; incubation temp 37°C ±0.3°C; permanent glass cuvettes
Photometry: 0–3.0 OD range; 13 wavelengths 340–800 nm
Automated pre-dilution: up to 100× (predefined factors)
Sample quality checks: Lipemia, hemolysis, icterus indices; clot detection and probe crash protection
Calibration & QC: Calibration stability up to 90 days (assay dependent); 2D barcode calibrator assignment; Shewhart (Levey-Jennings) and multi-rule (Westgard) QC
Installation & environmental requirementsDimensions & weight (analyzer with monitor): Weight max 450 kg; Length 2100 mm; Height 1602 mm; Depth 770 mm. Minimum clearances: rear 500 mm; left 500 mm; right 500 mm; front 635 mm.
Drain: max distance 10 m; max height 1.5 m from floor. Concentrated waste: 5.5 L/h (normal), 7.5 L/h HbA1c; diluted waste: 12 L/h.
Water: max consumption 20 L/h; deionized CAP type II or better, bacteria free; continuous flow; resistivity > 0.5 MΩ·cm filtered at 0.5 µm.
Power & IT: Power 200–240V, 60 Hz. Computer: Intel Core i7-6700, 32GB DDR4, 512GB SSD, Windows 10. Communications: RS232, Ethernet TCP/IP. Monitor: 21.5 in touchscreen (1920×1080). Operating environment: 18°C–32°C; humidity 20%–80% RH; altitude ≤2000 m; noise <60 dB.
Additional notes- Unique continuous loading and DxLAB racks reduce sample sorting and simplify workflow.
- Configurable with ISEs to support higher throughput during surges.
Technical specifications- Model: DxC 500 AU
- Analytical method: Fully automated random-access clinical chemistry with STAT capability
- Principles: Spectrophotometry, potentiometry; colorimetry, turbidimetry, latex agglutination, indirect ISE
- Menu capacity: >135 preprogrammed photometrics (including LIH + HbA1c and ISE)
- Throughput: 400 photometric tests/h (up to 800 with ISE); ISE only up to 600/h; HbA1c up to 65/h
- Sample capacity: 168 tubes; STAT carousel 22 samples
- Reagent carousel: 76 refrigerated positions; supports 15/30/60 mL bottles with barcoding
- Photometric range/wavelengths: 0–3.0 OD; 13 wavelengths 340–800 nm
- Reaction volume/time/temp: 90–350 µL; up to 8 min 37.5 s; 37°C ±0.3°C
- I/O & computer: RS232, Ethernet TCP/IP; Intel i7, 32GB, 512GB SSD, Windows 10; 21.5 in touchscreen