Fully automated immunochemistry analyzer DxC 500i
for clinical diagnosticwith ISEwith ion-selective electrode

Fully automated immunochemistry analyzer - DxC 500i - Beckman Coulter - for clinical diagnostic / with ISE / with ion-selective electrode
Fully automated immunochemistry analyzer - DxC 500i - Beckman Coulter - for clinical diagnostic / with ISE / with ion-selective electrode
Fully automated immunochemistry analyzer - DxC 500i - Beckman Coulter - for clinical diagnostic / with ISE / with ion-selective electrode - image - 2
Fully automated immunochemistry analyzer - DxC 500i - Beckman Coulter - for clinical diagnostic / with ISE / with ion-selective electrode - image - 3
Add to favorites
Compare this product

Characteristics

Operation
fully automated
Applications
for clinical diagnostic
Technology
magnetic bead-based, with ISE, with ion-selective electrode
Analysis mode
chemiluminescence, chemiluminescence immunoassay, with immunoturbidimetry dosing, colorimetric
Sample type
plasma, serum, whole blood, amniotic fluid, urine
Tested parameter
HbA1c
Sampling mode
random access
Configuration
integrated system
Other characteristics
with barcode reader, with touch screen, with clot detection, with LIS connectivity
Other measured parameters
with HbA1c analyzer
Throughput

400 p/h

Sample volume

Max.: 0.11 ml
(0.00372 US fl oz)

Min.: 0.001 ml
(0.000034 US fl oz)

Test storage capacity

168 unit

Description

Overview
The DxC 500i Clinical Analyzer provides integrated clinical chemistry and immunoassay testing in one random-access system, designed for low- to mid-volume laboratories to deliver consistent, commutable patient results with streamlined operation.

Key benefits
  • High availability and serviceability to minimize downtime
  • Commutable patient results across chemistry and immunoassay methods
  • Broad assay portfolio including extensive chemistry and immunoassay menus
  • Optimized workflow to reduce hands-on time and speed critical results

Standardized features
  • Shared reagents and consumables compatible with legacy Beckman Coulter analyzers for simplified inventory
  • Familiar user interface to reduce operator training
  • DxLAB sample rack for efficient sample handling
  • SimpleSolve onboard guidance for in-situ notifications and guided fixes

Operational performance
  • Time to first result: from 8 minutes (chemistry) and about 18 minutes for selected immunoassays (example: βhCG)
  • Throughput: up to 400 photometric chemistry tests/hour (up to 800/hour with ISE), up to 100 immunoassay tests/hour (one-step assays)
  • Random access with priority sample handling and rapid response during testing surges
  • One-tube testing for chemistry and immunoassay to reduce sample transfers

FlexMode operations
  • Independent-side maintenance: run tests on one module while servicing the other
  • Dynamic Sample Handler to prioritize immunoassay or chemistry based on sample urgency
  • 12-lane sample handler with continuous loading for streamlined processing

Technical specifications
  • Analytical method: Fully automated, random-access integrated system with chemistry and immunoassay modules
  • Analytical principles: Spectrophotometry, potentiometry, chemiluminescence
  • Assay types: Chemistry (endpoint, rate, fixed point, indirect ISE); Immunoassay (competitive, sandwich, antibody detection)
  • Chemistry techniques: Colorimetry, turbidimetry, latex agglutination, indirect ISE; Immunoassay: paramagnetic bead-based methods
  • Assay menu: >170 preprogrammed assays available
  • Throughput details: 400 photometric tests/hour (up to 800 with ISE); 100 immunoassay tests/hour (one-step); ISE max 600 tests/hour; HbA1c max 65 tests/hour
  • Sample types: Serum, plasma, urine, CSF, whole blood, amniotic fluid (assay dependent)
  • Sample capacity: 168 tubes (24 racks × 7) with continuous loading; Chemistry STAT carousel 22 refrigerated samples; Immunoassay 60 samples in 6 carousel racks
  • Barcoded reagents: automatic tracking of tests, expiration and lot numbers
  • Supported container sizes: Tubes 13×75 mm, 13×100 mm, 16×100 mm; Sarstedt S‑Monovette 4.9/5.5/7.5 mL; cups and inserts including pediatric options; auto-aliquot supported
  • Sample volumes: Chemistry 1.0–25 μL (0.1 μL increments); Immunoassay 10–110 μL
  • Barcode symbologies: Codabar (NW-7), CODE 39, CODE 128, ISBT-128, Interleaved 2 of 5
  • Compartment temperatures: chemistry reagent 4°C–12°C; cuvette 37°C ±0.3°C; immunoassay reagents 1°C–5°C; incubator ~36.4°C
  • Reagent capacity: Chemistry 76 refrigerated positions (15/30/60 mL); Immunoassay 24 refrigerated positions
  • Reaction vessels: Chemistry permanent glass cuvettes; Immunoassay disposable reaction vessels
  • Sample quality checks: lipemia/hemolysis/icterus indices, clot detection, probe crash protection
  • Calibration and QC: chemistry curves up to 90 days, immunoassay up to 63 days; Shewhart and multi-rule (Westgard) QC
  • Dimensions & weight: approx. 650 kg dry weight; L 254 cm × H 124 cm (160 cm with monitor) × D 77 cm
  • Clearance: rear 64 cm; left/right 50 cm; front 63.5 cm; top from chemistry top 124 cm
  • Waste: chemistry drain max 10 m/height 1.5 m; immunoassay drain max 3 m/height 0.9 m; specified waste flow rates
  • Water: max 20 L/hour, deionized CAP type II or better, bacteria-free, continuous flow
  • Power: 200–240 V (50/60 Hz), 3.5 kVA
  • Computer & display: 960 GB SSD, Windows 10, RS232/Ethernet LIS interfaces; 21.5" projected-capacitive touchscreen (1920×1080)
  • Operating environment: 18°C–28°C, 20%–80% RH (no condensation), altitude ≤2,000 m, noise <75 dBA
  • Communication: serial or TCP connection to LIS per ASTM protocol