●single channel counter
● samples test per hour
Product Description
Features:
● 3 DIFF scatter plots and 1 BASO scatter plot, 2 histograms are provided;
● 2 leukocyte test channel, basophils detected by a separate channel to ensure accuracy;
● Sheath flow digital fitting technology eliminates interference to ensure accuracy of RDW-CV and RDW-SD
● Support external barcode scanner
● Support laser, inkjet, dot matrix and other printers.
Specifications:
Test method: "Sheath flow technology and multi-angle laser scattering" realize WBC counting and classification;"Environmentally friendly cyanide-free reagent" realizes
HGB counting; "Impedance method" realizes RBC and PLT counting
Sample type: Venous, capillary, pre-diluted;
Accuracy: WBC CV≤2.0% (3.5~15.00×109 / L);
RBC CV≤1.5% (3.50~6.00×1012 / L);
HGB CV≤1.5% (110~180 g / L);
PLT CV≤6.0% (100~149×109 / L); CV≤4.0% (150~500×109 / L);
Linear range: WBC: 0~100×109/L; RBC: 0~8×1012/L; PLT: 0~1000×109/L; HGB: 0~250g/L;
Carry over: WBC≤0.5%; RBC≤0.5%; HGB≤0.6%; PLT≤1.0%;
QC: L-J, X-B, X-R, X, etc;
Calibration: Manual calibration, calibration material calibration, and fresh blood calibration
Reference: Male, Female, Children, Neonate
Display: 10.4 inches color touch screen
Storage: Up to 50,000 sample results+ all result scatter plots and histograms
Input / output: 4×USB, 1×LAN (support HL7 protocol)
Power supply: 100-240V±10%, 50/60Hz~±1Hz