ICP-AES spectrometer SPE51-1200
for the pharmaceutical industrydirect sample analysisfor the food industry

ICP-AES spectrometer - SPE51-1200 - labstac - for the pharmaceutical industry / direct sample analysis / for the food industry
ICP-AES spectrometer - SPE51-1200 - labstac - for the pharmaceutical industry / direct sample analysis / for the food industry
Add to favorites
Compare this product

Characteristics

Type
ICP-AES
Applications
for the pharmaceutical industry, for biomedical research, for research and development, for environmental analysis, for the food industry, for biochemistry, for metal analysis, direct sample analysis
Wavelength

Max.: 800 nm

Min.: 190 nm

Description

Application
  • Used in Environmental, Metallurgical, Geological, Petrochemical, Pharmaceutical, Food safety

Key Features
  • Petrochemical. Ion Beam Etching Holographic Grating, 3600 L/mm or 2400 L/mm.

Specifications
RFPower technical parameter:
      Circuit type: solid-state RF power supply, with function of automatch
      Frequency: 27.12 MHz ± 0.05%
      Frequency Stability: < 0.1 %
      Power Output: 800 W - 1200 W
      Power Output Stability: < 0.3 %
      Escaped RF radiation: 30 cm away from the instrument, electric field: E < 2V/m
Sampling System Technical Parameter:
      Output working coil inner diameter: 25 mm
      Torque tube: Three concentric, external diameter 20 mm
      Coaxial nebulizer: Outer diameter 6 mm
      Double barrel atomizing chamber: Outer diameter 34 mm
Gas Flow Controls:
      Plasma Argon Flowmeter: (100-1000) L/h (1.6-16 L/min)
      Auxiliary Argon Flowmeter: (10-100) L/h (0.16-1.66 L/min)
      Carrier Argon Flowmeter: (10-100) L/h (0.16-1.66 L/min)
      Pressure Maintaining Valve: 0 - 0.4 MPa
      Cooling Water: Temperature: 20-25 °C, Rate of Flow >5 L/min, Hydraulic Pressure >0.1 Mpa
Spectrometer:
      Optics: Czerny-Turner type
      Focal length: 1000 mm
      Grating: Ion Beam Etching Holographic Grating, 3600 L/mm or 2400 L/mm
      Reciprocal linear dispersion: 0.26 nm/mm
      Resolution: ≤ 0.007 nm (3600 line grating); ≤ 0.015 nm (2400 line grating)
      Wavelength range: 3600 line grating: (190 nm ~ 500) nm; 2400 line grating: (190 nm ~ 800) nm
      Minimum pace of stepping motor: ≤ 0.0006 nm
      Exit Slit: 12 µm
      Entrance Slit: 10 µm
Photoelectric Converter Performance:
      Photomultiplier tube specification: R293/R928
      Negative HV on PMT: 0 - 1000 V
      Stability: < 0.05 %

Technical Characteristics / Specifications
  • Model: SPE51-1200
  • Solid-state RF power supply with automatch function
  • Frequency: 27.12 MHz ± 0.05%
  • Power Output: 800 W - 1200 W
  • Optics: Czerny-Turner type
  • Grating: Ion Beam Etching Holographic Grating, 3600 L/mm or 2400 L/mm
  • Resolution: ≤ 0.007 nm (3600 line grating); ≤ 0.015 nm (2400 line grating)
  • Wavelength range: 190 nm ~ 500 nm (3600 line grating); 190 nm ~ 800 nm (2400 line grating)
  • Photomultiplier tube: R293/R928