OverviewThe Zetasizer Lab is an entry‑level dynamic light scattering (DLS) and zeta potential instrument for particle and molecular size analysis. It integrates Adaptive Correlation, M3‑PALS and a constant‑current zeta mode, and supports extended size‑range analysis when used with the ZSU1002 low‑volume capillary cell. ZS Xplorer software provides data quality feedback and guided analysis.
Key features- Classical 90° side‑scatter DLS for particle/molecular size (typical 0.3 nm–10 μm; up to 15 μm with extended analysis)
- M3‑PALS and constant‑current zeta mode for zeta potential and electrophoretic mobility
- Support for low sample volumes: size from 3 µL (capillary), zeta from 20 µL
- Static light scattering at 90° for molecular weight estimation
- On‑site upgrade path to Zetasizer Pro or Zetasizer Ultra
- Included ZS Xplorer analytics with instant data‑quality feedback
Model variants- Blue Label — recommended for routine colloids (metallic colloids, metal oxides, pigments)
- Red Label — recommended for challenging samples (proteins, surfactant solutions, low solid content)
Which Zetasizer is right for me?Comparison (overview): Zetasizer Ultra | Zetasizer Pro | Zetasizer Lab — Size principle: NIBS & MADLS | NIBS | Classical 90° DLS; Zeta principle: ELS with M3‑PALS & constant current (all models); Low‑volume capillary support: Yes | No | Yes.
Key applications- Colloidal silica — reproducible sizing for quality control
- Titanium dioxide — aggregation/agglomeration and stability vs. pH or salt concentration
- Ceramics — particle size distributions and multimodal populations
- Proteins — low‑volume characterization using capillary cell
- Fluorescent nanomaterials (C‑dots) — measurement considerations for fluorescence
- Viral vectors (AAV) — size distribution and indicative titer (MADLS on Ultra for concentration/titer)
Technical specifications- Measurement principle (size): Classical 90° DLS; measurement angles: 90°, 13°
- Size range: 0.3 nm – 15 μm (typical 0.3 nm–10 μm)
- Minimum sample volume (size): 3 µL (capillary)
- Zeta range: 3.8 nm – 100 μm; minimum volume (zeta): 20 µL; sample conductivity up to 260 mS/cm
- Static light scattering (molecular weight): Red Label 300 Da–20 MDa; Blue Label 980 Da–20 MDa (±10%)
- Detector: Avalanche photodiode; Laser: He‑Ne 633 nm (Blue 4 mW / Red 10 mW)
- Temperature control: 0 °C – 120 °C (dry N2 purge required below dewpoint)
- Dimensions: 322 × 565 × 245 mm; Weight: 19 kg; Power: AC 100–240 V, 50–60 Hz
- Standards: ISO22412:2025; ISO13099‑2:2025; relevant USP and ASTM methods