Technical specifications
(1) Defibrillator type: Manual, Synchronized, A synchronized
(2) Output waveform: Monophonic technology /Biphasic technology (optional)
(3) Energy accuracy: 1%
(4) Energy selection: 0,3,5,7,10,20,30,50,100,200,300,360 joule(nominal at a resistance of 50)
(5) Charging time: Maximum 8 seconds to 360 Joules
(6) Standard Adult/Pediatric Paddles: Reusable external adult paddles (pediatric paddles integrated)
Power requirements
(a) AC Power supply: 110V, 60 Hz /220V, 50 Hz
(b) Vehicle Voltage: DC 12V
Internal Battery Backup
(1) Battery type: Rechargeable lead-acid battery/ lithium battery (optional)
(2) Charging time: Time-Minimum of 4 hours for full charging
(3) Battery charging: During AC Power Supply operation of the device.
(4) Charging /Low Battery Indication: Red LED
(5) Battery Capacity: 100% of the battery will provide 120 min of monitoring; and 30 discharges at 360 joules.
Clear Button
(1) CLEAR BUTTON Clear the energy selection and discharge the energy by inside circuit.
(2) CHARGE To indicate charging process for the defibrillator battery. The lamp went out if full capacity.
(3) BATTERY INDICATOR To show the DC work of inner battery.
(4) DEFIBRILLATING ENERGY INDICATOR
(5) ENERGY SELECTOR To select the energy of the heart defibrillating treatment.