Product DescriptionThe AV-3010 is a mechanical ventilator intended for ICU respiratory support for adult, pediatric and infant patients. It regulates breathing rate, inspiratory and expiratory phases and supports humidification. Core components are selected for clinical reliability and the device integrates a 12.1-inch colour touch LCD for bedside operation.
Features- 12.1-inch colour touch screen with configurable display
- Simple interface for accurate control and intuitive operation
- Applicable for adult, pediatric and infant patients
- Multi-mode ventilation: VCV, SIMV, SIMV/PSV, PSV, PCV, APRV, APNEA, CPAP, BIPAP, MANUAL, SIGH, NIV
- Supports invasive and non-invasive ventilation
- Integrated air compressor and pipeline gas inputs (O2, Air)
- Self-test and calibration; auto-clean exhaust valve
- Support for nebulization and high-flow oxygen (up to 60 L/min)
Technical SpecificationsPhysical:
Dimensions: 45 × 55 × 140 cm
Weight: 65 kg
Screen: 12.1-inch colour touch LCD
Pipeline supply:
Gas inputs: O2 and medical air (pipeline input range 280–600 MPa)
Air compressor:
Power: AC 220V ±10%, 50Hz ±2%
Compressor output pressure: 0.4 ±0.1 MPa
Noise: ≤50 dB
Battery:
Type: Lead-acid 12V, 4Ah
Typical runtime: ~180 minutes+
Ventilation parameters:
Modes: VCV, APNEA, SIMV, SIMV/PSV, MANUAL, CPAP, PSV, PCV, BIPAP, SIGH, APRV, NIV
Tidal volume: 20–2000 mL
Pressure range: 5–80 cmH2O
Respiration rate: 1–100 bpm (1–40 bpm in SIMV)
Inspiration time: 0.2–12 s
FiO2: 21%–100%
Triggers: Ptrigger -20–0 cmH2O; Ftrigger 0.5–30 L
Monitoring & alarms:
Monitoring: Vti/Vte, Ppeak, Pmean, Pplateau, PEEP, NIF, flow and respiratory indices; CO2 optional
Alarms: visual and audio alarms for tidal volume, minute volume, Paw, FiO2, oxygen limitation, air/power failures
Model and key specifications- Model: AV-3010
- Display: 12.1-inch colour touch LCD
- Dimensions: 45 × 55 × 140 cm; Weight: 65 kg
- Power: AC 220V ±10%, 50Hz ±2%; Battery: 12V 4Ah (~180 min)
- Tidal volume: 20–2000 mL; Pressure: 5–80 cmH2O; RR: 1–100 bpm
- Gas inputs: O2 and medical air; Compressor output: 0.4 ±0.1 MPa
- Optional: auxiliary suction, CO2 monitoring, nebulization