The TD-1090 Temperature Chamber is designed for accuracy verification of ear and forehead thermometers. It features an adjustable target temperature in both °C and °F, and supports auto-recalibration to the TD-1107 thermometer. The device offers data export capability and simple operation for ease of use.
Main Features- Accuracy verification for ear and forehead thermometers
- Adjustable target temperature in °C & °F
- Auto-recalibration to TD-1107 thermometer
- Data export capability
- Simple operation
Specifications- Display Resolution: 0.1°C/0.1°F
- Power & Fuse(s): 100 ~ 240VAC (50/60Hz) / 3A
- Memory: 1000
- Operating Condition: 16°C to 32°C (60.8°F to 89.6°F), 85% RH or less
- Temperature Range (0.98 emissivity): 37.0°C ~ 41.0°C (98.6°F ~ 105.8°F)
- Acceptable Recalibration Range: ±0.5°C (±0.9°F)
- Stability: ±0.03°C at 37°C
- Measurement Units: °C (default) or °F
- Warm-up Time at 25°C Ambient: 30 minutes
- Nominal Emissivity: 0.98 emissivity
- Computer Interface: USB HID
- Dimension: 300mm x 350mm x 160mm
- Units per Carton: 1
- Weight per Carton: 6.93 kg
- Carton size: 41.3 (L) x 22.3 (W) x 37.6 (H) cm