Product overviewThe HTY-AW01-TH water activity meter determines sample aw using the mirror condensation method with controlled temperature and humidity. Suitable for quality control and R&D in pharmaceuticals, food, cosmetics and environmental testing to assess microbial risk and shelf life.
Key features- Mirror condensation method: Traceable to SI units and aligned with USP recommended methods (USP 922, USP 1112).
- High accuracy: Resolution 0.0001 aw and high-precision capability (±0.003 aw) with a maximum allowable error ±0.02 aw.
- Active temperature control: 10–50 °C range with ±0.2 °C accuracy for fast and repeatable measurements (typical test time ≤30 min; with active control measurement time can be reduced to ~4–6 min).
- Data management: 3-level user access, audit trail and 21 CFR Part 11 support; long-term storage (>10 years).
- Low operating cost: Inexpensive consumables and minimal sensor maintenance.
Specifications- Test range: 0–1.000 aw
- Max. allowable error: ±0.02 aw
- High accuracy capability: ±0.003 aw
- Resolution: 0.0001 aw
- Repeatability error: ≤5%
- Temperature control range: 10–50 °C; accuracy ±0.2 °C; sample resolution 0.01 °C
- Environment: 4–50 °C; 10%–85% RH (non-condensing)
- Sample cup capacity: 7 mL
- Input power: 100–240 VAC ±10%, 50/60 Hz; power ≤100 W
- Data storage: >10 years (records aw, temperature, time, date, operator, sensor type)
Compliance- USP 922 Water Activity
- USP 1112 Application of Water Activity Determination in Non-sterile Drugs
- USP 1241 Water–Solid Interactions in Drug Systems
- 21 CFR Part 11 compliant data handling
ApplicationsDesigned for routine QC and R&D testing of raw materials and finished products across regulated industries.
- Pharmaceuticals
- Food and beverages
- Cosmetics
- Biodegradable materials, soil and compost analysis
Technical details- Model: HTY-AW01-TH
- Sample cup capacity: 7 mL
- Typical test time: ≤30 min (can be reduced to ~4–6 min with active temperature control)
- Supports audit trail and multi-user access (3 levels)