Product OverviewThe basic control system is a 7" color touch screen interface for fume hood control, providing core control and reminder functions.
Key features- Lighting and power control outlets
- Fan ON / OFF
- Periodic test reminder
- Clock and date setup
Models and references- Basic — Part number: FH-CON; Screen: 7"; Airflow control: Manual air velocity control; Air velocity range: N/A; Low alarm set point: N/A; Outputs: N/A; Analog in: N/A
- AFA — Part number: FH-AFA; Screen: 7"; Airflow control: Manual air velocity control with airflow alarm; Air velocity range: 0-2 m/s (0-400 FPM); Low alarm set point: 0.35 m/s (70 FPM); Outputs: N/A; Analog in: N/A
- VAV — Part number: FH-VAV; Screen: 10.1"; Airflow control: Automatic air velocity adjustment; Air velocity range: 0-2 m/s (0-400 FPM); Low alarm set point: 0.35 m/s (70 FPM); Output: 0-10 VDC; Analog in: 0-5 VDC
- Active + VAV — Part number: FH-ACT; Screen: 10.1"; Airflow control: Automatic sash motion with proximity sensing + automatic air velocity adjustment; Air velocity range: 0-2 m/s (0-400 FPM); Low alarm set point: 0.35 m/s (70 FPM); Output: 0-10 VDC; Analog in: 0-5 VDC
Technical specifications- Brand: Topair Systems
- Control interface: 7" color touch screen (Basic, AFA) and 10.1" option (VAV, Active + VAV)
- Available control modes: manual air velocity control; manual with airflow alarm; automatic air velocity adjustment; automatic sash motion with proximity sensing (Active + VAV)
- Air velocity range (AFA / VAV / Active + VAV): 0-2 m/s (0-400 FPM)
- Low alarm set point (where applicable): 0.35 m/s (70 FPM)
- Analog outputs (where applicable): 0-10 VDC; Analog inputs: 0-5 VDC
- Part numbers: FH-CON (Basic), FH-AFA (AFA), FH-VAV (VAV), FH-ACT (Active + VAV)
- Input power requirement: 110-230 V, 50/60 Hz
Additional informationInput Power: 110-230 V 50/60 Hz