10.1" touchscreen operator interface for real-time monitoring and configuration
Automatic air velocity control with installed and included VFD on VAV variants
Temperature and humidity display for environmental monitoring
Lighting and power outlet control outputs
Fan ON/OFF control and integrated scheduled test reminders
Clock and date configuration
Red light alarm and configurable airflow alarms (where specified)
Sash motion automation with proximity sensing on Active + VAV model
Models / Technical table
Basic — FH-CON — 7" screen — Air velocity manual control — Airflow control type: manual — Air velocity range: N/A — Low alarm range set point: N/A — Output: N/A — Analog in: N/A
AFA — FH-AFA — 7" screen — Air velocity manual control with airflow alarm — Air velocity range: 0-2 m/s (0-400 FPM) — Low alarm set point: 0.35 m/s (70 FPM) — Output: N/A — Analog in: N/A
VAV — FH-VAV — 10.1" screen — Air velocity automatic adjustment — Air velocity range: 0-2 m/s (0-400 FPM) — Low alarm set point: 0.35 m/s (70 FPM) — Output: 0-10 VDC — Analog in: 0-5 VDC
Active + VAV — FH-ACT — 10.1" screen — Automatic sash motion with proximity sensing + Air velocity automatic adjustment — Air velocity range: 0-2 m/s (0-400 FPM) — Low alarm set point: 0.35 m/s (70 FPM) — Output: 0-10 VDC — Analog in: 0-5 VDC
Features (summary)
Operator interface with clear 10.1" display for system status and controls
Built-in VFD for automated fan/air velocity control on VAV models
Environmental readouts: temperature and humidity
Integrated controls for lighting, power outlets and fan operations
Alarm management: red light alarm and configurable airflow alarms
Periodic test reminders to support maintenance schedules
Date and time configuration for event logging and scheduling
Technical specifications
Input power: 110-230 V, 50/60 Hz
Available variants and part numbers: FH-CON (Basic), FH-AFA (AFA), FH-VAV (VAV), FH-ACT (Active + VAV)
Screen sizes: 7" (Basic, AFA) and 10.1" (VAV, Active + VAV)
Air velocity range (VAV variants): 0-2 m/s (0-400 FPM)
Typical low alarm set point: 0.35 m/s (70 FPM)
Analog outputs (VAV / Active + VAV): 0-10 VDC
Analog inputs (VAV / Active + VAV): 0-5 VDC
Sash motion automation (Active + VAV): automatic sash movement with proximity sensing
Includes installed VFD for automatic air velocity control on VAV variants