video corpo
  • Products
  • Catalogs
  • News & Trends
  • Exhibitions

Pressure tester BPA700
flow rateleak-proofingfor blood pressure monitors

Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors
Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors
Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors - image - 2
Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors - image - 3
Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors - image - 4
Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors - image - 5
Pressure tester - BPA700 - WhaleTeq Co., Ltd - flow rate / leak-proofing / for blood pressure monitors - image - 6
Add to favorites
Compare this product

Characteristics

Tested parameter
pressure, flow rate, leak-proofing
Applications
for blood pressure monitors
Configuration
benchtop, tabletop
Other characteristics
digital, USB, non-invasive
Weight

4.5 kg
(9.9 lb)

Width

31.5 cm
(12.4 in)

Height

8.8 cm
(3.5 in)

Description

Overview
The BPA700 is an NIBP simulator and analyzer designed for preclinical verification of non-invasive blood pressure monitors and algorithms. It addresses limitations of clinical-only testing by enabling recording and playback of blood pressure raw data, allowing engineers to validate algorithms with real-case signals and reuse datasets for iterative development and verification.

Key features
  • Static pressure test — verifies blood pressure reading accuracy and cuff performance.
  • Blood pressure test — simulates systolic and diastolic pressures with customizable pulse envelope.
  • Leak test — measures leakage rate to assess cuff and system tightness.
  • Overpressure test — validates safety mechanism response.
  • Record and play data — records blood pressure raw signals and converts them to playback files for algorithm validation.
  • Standard Assistant Software — integrates IEC 80601-2-30 and YY9706.230 requirements into 14 test items for simplified standard-compliance testing.
  • Standalone and automated modes — supports manual tests and Auto Sequence for automated test flows without coding; SDKs available for custom automation.


Highlights
  • Flexible pulse envelope settings to reproduce varied hemodynamic conditions.
  • SDKs for developing automated testing programs and integration with test benches.
  • Auto Sequence function for creating automated test sequences without programming.
  • ISO/TS 81060-5 compliant NIBP simulator.
  • Optional Standard Assistant Software simplifies IEC 80601-2-30 and YY9706.230 compliance and exports test reports.
  • Wide dynamic and static pressure ranges to simulate diverse blood pressure levels.


Validation using raw data playback
  • Raw data playback enables direct testing of NIBP algorithms with clinical-like signals, reducing dependence on costly clinical trials.
  • Raw data recording builds reusable databases for future validation and regression testing.


Ordering / Package contents (summary)
Model No.: BPA700. Model name: NIBP simulator and analyzer. Package contents include USB transfer cable (Type-A to Type-B, 1.8 m), two tubes (14 cm) and one T-connector.

Documentation and tools (summary)
Software, datasheet, user manual and SDK are provided as downloadable resources (software packages and documentation available separately).

Technical specifications
  • Pressure unit: mmHg, kPa
  • Manometer: range 0 to 400 mmHg; resolution 0.1 mmHg; accuracy ±(0.3% of reading + 0.5 mmHg)
  • Pressure source: target pressure 20 to 400 mmHg; settle time 5 s; resolution 1 mmHg; accuracy ±0.5 mmHg
  • Pulse rate: range 30 to 300 bpm; accuracy ±1 bpm; pulse amplitude up to 2 mmHg (500 ml air reservoir); pulse volume 0 to 2 ml
  • Waveform: oscillometric support; sample rate 200 Hz (5 ms)
  • Leak test: time 0 to 999 s; target pressure 20 to 400 mmHg; range 0 to 300 mmHg/min
  • Overpressure test: release time 0 to 999 s; automatic inflation; instantaneous 0 to 400 mmHg
  • BP envelope shift: systolic ±20 mmHg; diastolic ±20 mmHg
  • BP dynamic range: diastolic 10 to 250 mmHg; systolic 25 to 300 mmHg
  • Dynamic NIBP simulation repeatability: ±2 mmHg; standard deviation 0.05 mmHg
  • Self-leakage: <1 mmHg/min (500 ml air reservoir)
  • Environment: operating temperature 10°C to 40°C; storage 0°C to 50°C; humidity 0–90% RH (non-condensing)
  • Dimensions / display / weight: 326.4 × 315 × 88 mm; graphic LCD 320 × 240 pixels; weight 4.5 kg
  • Communication: USB
  • Power: AC 100–240 V, 50/60 Hz

VIDEO