OverviewCompact neutral electrode contact impedance tester designed for compliance with IEC 60601-2-2 (201.15.101.6). Ideal for medical equipment testing labs and compliance verification.
Features- Designed to meet IEC 60601-2-2:2009 (201.15.101.6) requirements for neutral electrode contact impedance testing
- Outputs a 200 mA signal with selectable added impedance from 0 to 50 Ω across 50 kHz–5 MHz
- Stable frequency generation using DDS (Direct Digital Synthesis)
- Built-in current monitoring port (BNC) — no separate expensive probe required
SpecificationsFrequency range: 50 kHz – 5 MHz
Maximum voltage: 11 Vrms (~31 Vpp) at ≥50 Ω load
Maximum current: >200 mArms at ≤50 Ω load
Output waveform: Sine
Frequency response: <±0.8 dB (50 Ω)
THD: <0.3% (-50.4 dB) at 200 mArms, 50 Ω, 1 MHz
Impedance: <3.2 Ω + j0.4 µH
Oscilloscope pairing: Use oscilloscope (Ch1: voltage probe; Ch2: HFPA150 current monitor port) to measure voltage and current
Current measurement accuracy: +10% / -0% (unilateral tolerance)
Current measurement frequency range: 50 kHz – 5 MHz
Current measurement port: BNC
Power supply: Adapter input 100–240 V, 50/60 Hz; output 12 V DC / 2 A
Environment: 10 °C – 40 °C, 30% – 90% RH (design conditions, not test values)
Dimensions / Weight: 18 x 18 x 4.8 cm (L x W x H) / 0.7 kg (net)
Ordering informationPart number: 100-HF00103
Model: HFPA150
Product name: NE contact impedance tester
Quantity: 1
Optional accessoriesPart number: 300-HFPA15D — Description: Calibration load for HFPA150 (1 Ω / 50 Ω) — Quantity: 1
Calibration service & extended warrantyPart number: YY0007 — Model: C3 — Description: Manufacturer 3-year calibration service (recommended annual calibration to verify performance)
Part number: YY0008 — Model: R3 — Description: Warranty extension from 1 year to 3 years
DocumentationDatasheet: HFPA150 datasheet
Manual: HFPA150 user manual (available for download)
Technical features / specifications- Model: HFPA150
- Frequency range: 50 kHz – 5 MHz
- Output current: >200 mArms (≤50 Ω load)
- Output voltage: 11 Vrms (~31 Vpp) (≥50 Ω load)
- Output waveform: Sine
- THD: <0.3% (200 mArms, 50 Ω, 1 MHz)
- Impedance: <3.2 Ω + j0.4 µH
- Current measurement port: BNC
- Power: 12 V DC / 2 A (input 100–240 V, 50/60 Hz)
- Operating environment: 10 °C–40 °C, 30%–90% RH
- Dimensions / Weight: 18 x 18 x 4.8 cm / 0.7 kg