Flexible sample management:continous sample loading by racks. STAT priority and rerun.
HbA1c Test with Whole Blood: onboard hemolysis without sample preparation.
Cost efficiency: dedicated reagents, low reagents consumption. Effective cuvettes washing. Long shelf life of electrodes.
Integrated ISE Module: 1 blood tube for electrolytes and photometric chemistry tests. Automatic calibration.
Advanced software. Touch-screen monitor & Three axis swing arm
Small footprint: 1.6 m2
Optional automatic tube unclogging function
Possibility of integration with CLIA 2600 immunoassay module
Main characteristics:
Speed:
1000 photometric tests/h
300 ISE tests/h
1200 tests/h with ISE
Measurement system:
Reaction system:
Reaction volume: 80 – 280 µL
Operating temperature: 37 ºC
Light source: tungsten halogen lamp
Diffraction grating: light dispersion
Wavelength: 16 wavelengths (340 nm, 380 nm, 412 nm, 450 nm, 480 nm, 505 nm, 546 nm, 570 nm, 605 nm, 630 nm, 660 nm, 700 nm, 740 nm, 770 nm, 800 nm, 850 nm)
Absorbance range: 0~3.5 Abs (10 mm conversion)
Electrolyte module:
Principle: Direct potentiometry
Ion-selective electrode: sodium, potassium, and chloride
Control and calibration:
Calibration mode: linear (one-point, two-point, and multi-point), logit-log 4P, logit-log 5P, spline, exponential, polynomial, parabola
Control standards: multiple Westgard rules, Levy-Jennings plots, twin plot
Sampling system
Samples:
180 samples in continuous sample loading racks, STAT priority
Sample volume: 1 - 25 µL, with 0.1 µL increments
Sample needle: liquid level detection, clot detection, and collision protection