With true family commonality and enhanced system and assay dynamics, the ARCHITECT family allows for integration without compromise.
Equivalent patient results
Identical technology, reagents and assay protocols generate patient results that are equivalent from system to system.
Identical software
ARCHITECT offers identical software across the entire family of instrumentation to streamline training and enhance ease of use.
Universal architect sample carrier
A universal sample carrier and common reagent cartridges are utilized across ARCHITECT systems.
The ARCHITECT i1000SR immunoassay analyzer meets your laboratory’s high standards by delivering STAT results when you need them.
Sample Tubes -
Height: 72-102 mm
Diameter: 9.6-16.1 mm
Sample Cup - Yes (50 µl dead volume)
Sample Capacity - 65
Sample Barcode Types - Code 39, Codabar, Interleaved 2 of 5, Code 128
Sample Result Storage - 50,000
Automatic Dilution - Yes
Sample Probe Carryover - ≤0.1 parts per million
Reagent Capacity - 25 refrigerated positions
Reagent Type - 100% liquid ready-to-use
Reagent Onboard Stability - 14-30 days
Calibration Frequency - Calibrate with new lot number, if controls are out of range or if specified otherwise within the package insert
Sample, Clot and Bubble Detection - Yes
Reagent Pressure Monitoring - Yes
Sample Interference Measurement - No
System Control Center - 1 SCC, with color touchscreen monitor, keyboard and mouse
Onboard Maintenance Records - Yes
Online Error Code Help - Yes
Host Interface - Bidirectional, serial RS-232 interface, host query option available