Analysis of complex, multiple-pulse waveforms
Standardized industrial HF measurement
HF leakage current measurement
REM/ARM/CQM test adjustable in 1 O steps
Ultra-fast digitizing of HF waveforms
Sampling of up to 16 million data points
Internal precision loads adjustable from 1 to 6400 O in 1 O steps
Integrated real-time operating system with ¼ VGA color touch-screen
Automated, freely definable test sequences
Automated current-power curves with various power levels per load value setting
Compatible with external test loads
Definable screen display
USB, RS 232 and Ethernet communication ports
Dedicated connectors for external keyboard and mouse
Automatic or manual activation of the ESU generator during the load curve test
Remote control of the ESU generator is possible
Generation of test reports
Feature Description
The SECULIFE ES PRIME offers an unparalleled sampling rate of 16 million data points. This permits extremely stable measurement of the output waveforms of the latest generation of HF generators, which make use of multiple-pulse waveforms. The internal, low-impedance resistance decade has a range of 0 to 6400 O and can be adjusted in 1 O steps. The device is controlled by internal processors and includes an ideally matched combination of hardware and software. This permits highly precise, accurate measurement, even of HF waveforms with excessive interference, for example like a spray function. Each data point is analyzed and delivers highly precise measurement results.
The SECULIFE ES PRIME is equipped with internal relays for controlling the test circuit.