OverviewControlled CO2 incubation for reproducible cell culture workflows. Precise temperature, CO2 and humidity management, hygienic design and networked control suited to laboratory and clinical environments.
Range of functionsCalibration: Three freely selectable setpoints: temperature, humidity and CO2
Interfaces: USB‑C; Ethernet LAN
Timer: Immediate start; SetpointWAIT (process time starts after all setpoints reached); Timer on Event (starts at defined time)
Language options: English (default), German, French, Spanish, Portuguese, Chinese, Arabic, Hindi, Polish, Czech, Hungarian, Russian, Italian, Serbian
Device functions: Date/time (time zones, summer/winter time); Display brightness; DistanceView (1–4 selectable parameters); Network settings (IP, subnet, DHCP); Language and buzzer settings; Power‑failure behaviour; Timer functions
Adjustable parameters: Temperature; CO2; Humidity (option K7); O2 (option T6)
Control unit: 6.86" IPS multi‑touch display (480 × 1280 px) showing actual/setpoint values and parameter curves.
O2 (with option T6)Control: Anaerobic O2 control via filtered N2 inlet. Zirconium‑oxide sensor with monitoring and error display.
Adjustment precision: 0.1 %
Setting range: 1–20 % O2
TemperatureHeating: Large-area multi-zone all‑round heating with additional door, seal and rear heating
Setting range: +20 to +50 °C
Operating range: from 5 °C above ambient up to +50 °C
Sterilisation programme: 60 min at 180 °C (sensors may remain installed)
Setting accuracy: 0.1 °C
Spatial variation at 37 °C: ±0.3 K
Temporal fluctuation at 37 °C: ±0.1 K
HumidityHumidification: Passive water tank with condensate return; Peltier element for humidity limitation. Option K7: active closed‑loop control with capacitive sensor. Humidification via self‑priming micro‑dosing pump with hot steam; dehumidification via diaphragm pump.
Accuracy (K7): 0.5 % RH
Setting range: 40–98 % RH and rh‑Off
CO2Control: Dual‑beam NDIR CO2 sensor with integrated monitoring and pressure‑compensated measurement. Filtered CO2 supply.
Temporal fluctuation: ±0.2 %
Setting range: 0–20 % CO2
SafetyPower failure modes: Manual and program behaviours with retained parameters and defined continuation/query logic depending on duration
Temperature monitoring: Dual Pt1000 sensors DIN Class A with mutual monitoring and automatic takeover on fault
Alarms: visual and acoustic
Standard equipmentStandard option: Door‑open detection (V5)
Factory calibration certificates: various options (Standard / K7 / T6) for +37 °C, 5 % CO2, 90 % RH and O2 where applicable
Accessories: CO2 connection set, hoses and clamps; water tank (B38737); stored sterilisation programme 60 min at +180 °C
Door: Insulated stainless steel door with 2‑point compression lock
Inner door: Glass inner door with Ø 8 mm sampling opening
Internals: 1 perforated stainless steel shelf (max. 5 positions)
Stainless steel interiorInternal dimensions (W × H × D): 560 × 480 × 400 mm
Inner cover: 37.5 mm
Volume: 108 L
Max. internals: 5
Max. chamber load: 75 kg
Max. load per internal: 15 kg
Textured stainless steel casingExternal dimensions (W × H × D): 708 × 683 × 589 mm
Installation: 4 adjustable feet
Electrical data230 V, 50/60 Hz — approx. 1300 W
115 V, 50/60 Hz — approx. 1300 W
Ambient conditionsClearances: Rear ≥ 15 cm; Top ≥ 20 cm; Sides ≥ 5 cm
Ambient temperature: 10–35 °C
Max. installation altitude: 2000 m
Overvoltage category: II
Pollution degree: 2
Packing / shipping dataTransport: Unit must be transported upright
Customs tariff number: 8419 8998
Country of origin: Germany
WEEE‑Reg. No.: DE 66812464
Packaging dimensions (W × H × D): 825 × 967 × 815 mm
Net weight: approx. 70 kg
Gross weight: approx. 85 kg
CertificationsStandard units are safety‑approved and bear the CE mark
Technical specifications- Brand: Memmert
- Model: ICO105
- Product type: CO2 incubator
- Internal volume: 108 L
- Internal dimensions (W × H × D): 560 × 480 × 400 mm
- External dimensions (W × H × D): 708 × 683 × 589 mm
- Temperature setting range: +20 to +50 °C
- Temperature accuracy: 0.1 °C
- Temperature variation at 37 °C: ±0.3 K
- CO2 setting range: 0–20 %; temporal fluctuation ±0.2 %
- Humidity setting range: 40–98 % RH (rh‑Off); accuracy 0.5 % RH (option K7)
- O2 control (option T6): 1–20 %, precision 0.1 %
- Control/display: 6.86" IPS multi‑touch (480 × 1280 px)
- Power: 230/115 V, ≈1300 W
- Net weight: ≈70 kg; max. chamber load 75 kg
- Standard shelf: 1 perforated stainless steel shelf (max. 5 positions)
- Country of origin: Germany