OverviewControlled CO2 incubation for cell and tissue culture with precise humidity management, hygienic chamber design and laboratory connectivity options.
Range of functions- Calibration: three freely selectable setpoints – temperature, humidity, CO2
- Interfaces: USB-C, Ethernet LAN
- Timer modes: immediate start; SetpointWAIT (starts after reaching all setpoints); Timer on Event (starts at a defined time)
- Language settings: English (default), German, French, Spanish, Portuguese, Chinese, Arabic, Hindi, Polish, Czech, Hungarian, Russian, Italian, Serbian
Appliance functionsDate and time (time zones, summer/winter time); display brightness; DistanceView (1–4 selectable parameters, adjustable activation time); network settings (IP, subnet, DHCP); language and buzzer/keystroke tone settings; power-failure behaviour; timer functions.
Adjustable parameters- Temperature
- CO2
- Humidity (requires option K7)
- O2 (requires option T6)
Control unit6.86" IPS multi-touch display (480 × 1280 px) for graphical presentation of actual/setpoint values and parameter curves.
O2 (only with T6 option)Anaerobic O2 control via nitrogen inlet. Zirconium oxide sensor with integrated monitoring and error display. Filtered N2 supply. Adjustment precision: 0.1 %. Setting range: 1–20 % O2.
TemperatureLarge-area multi-zone all-round heating with additional door, seal and rear-wall heating. Setting range: +20 to +50 °C. Operational range: from 5 °C above ambient up to +50 °C. Built-in sterilisation programme: 60 min at 180 °C (sensors can remain in place). Setting accuracy: 0.1 °C. Spatial variation at +37 °C: ±0.3 K. Temporal fluctuation at 37 °C: ±0.1 K.
HumidityPassive humidification via integrated water tank with condensate return; humidity limitation by Peltier element. Option K7: active closed-loop humidity control with capacitive sensor, micro-dosing pump for hot-steam humidification and diaphragm pump dehumidification. Accuracy: 0.5 % RH. Setting range: 40–98 % RH and rh‑Off.
CO2Dual-beam NDIR CO2 control with pressure-compensated measurement and integrated monitoring. Filtered CO2 supply. Temporal fluctuation: ±0.2 %. Setting range: 0–20 % CO2.
SafetyPower-failure behaviour: manual and program modes with differentiated handling depending on outage duration; parameters remain stored. Temperature monitoring by dual Pt1000 sensors (DIN Class A) with mutual monitoring and automatic takeover. Visual and acoustic alarms.
Standard equipmentStandard option: door-open detection (V5). Factory calibration certificates available (Standard / Option K7 / Option T6). Included accessories: CO2 connection set, hose with coupling and clamp; water tank B38737; water tank with connection hose; stored sterilisation programme 60 min at +180 °C. Door: fully insulated stainless-steel door with 2-point compression lock. Inner door: glass with 8 mm Ø port for gas sampling. Internals: 2 perforated stainless-steel shelves (max. 9).
Stainless steel interiorDimensions (W × H × D): 600 × 810 × 500 mm. Inner cover: 37.5 mm. Volume: 243 L. Max. number of internals: 9. Max. chamber load: 135 kg (15 kg per internal).
Textured stainless steel casingDimensions (W × H × D): 748 × 1013 × 689 mm. Installation: 4 adjustable feet.
Electrical data230 V, 50/60 Hz — approx. 1650 W. 115 V, 50/60 Hz — approx. 1650 W.
Ambient conditionsMinimum clearances: rear ≥15 cm, top ≥20 cm, sides ≥5 cm. Ambient temperature: 10 °C to 35 °C. Max. installation altitude: 2000 m. Overvoltage category II; pollution degree 2.
Packing / shipping dataTransport upright only. Customs tariff number: 8419 8998. Country of origin: Federal Republic of Germany. WEEE Reg. No.: DE 66812464. Dimensions incl. carton (W × H × D): 925 × 1277 × 854 mm. Net weight: approx. 113 kg. Gross weight incl. carton: approx. 132 kg.
Technical specifications- Model: ICO240
- Brand: Memmert
- Volume: 243 L
- Internal shelves: 2 perforated stainless-steel shelves (max. 9)
- Chamber dimensions (W × H × D): 600 × 810 × 500 mm
- Overall dimensions (W × H × D): 748 × 1013 × 689 mm
- Temperature setting range: +20 to +50 °C (operation from 5 °C above ambient)
- Temperature setting accuracy: 0.1 °C
- CO2 setting range: 0–20 % (temporal fluctuation ±0.2 %)
- Humidity setting range: 40–98 % RH (or rh‑Off); accuracy 0.5 % RH
- Power: approx. 1650 W
- Supply voltages: 230 V or 115 V, 50/60 Hz
- Max. chamber load: 135 kg (15 kg per internal)
- Net / gross weight: approx. 113 kg / 132 kg
- Country of origin: Federal Republic of Germany
- Customs tariff number: 8419 8998
- WEEE Reg. No.: DE 66812464