Exhaust fume hood AllerGard™ ES
containmentlaboratorybiological

exhaust fume hood
exhaust fume hood
exhaust fume hood
exhaust fume hood
exhaust fume hood
exhaust fume hood
exhaust fume hood
exhaust fume hood
exhaust fume hood
Add to favorites
Compare this product
 

Characteristics

Function
exhaust, containment
Applications
laboratory, biological, animal transfer, cage changing
Configuration
floor-standing
Filter type
HEPA, activated carbon
Air flow type
laminar flow
Exhaust mode
ductless
Other characteristics
stainless steel
Depth

777 mm, 778 mm
(30.59 in, 30.63 in)

Height

2,350 mm
(92.52 in)

Description

The NU-620 employs a “piston of air” design that is configurable with dual-sided access for workflows needing flexibility in tight spaces. Overview The AllerGard NU-620 Dual Access Animal Transfer Station (ATS) employs a “push-pull” pattern of airflow to provide clean air in the work zone for small animal cage caging and generate suction through air slots on the work surface for allergen containment. A supply impeller creates the “push” in the transfer station by drawing air through a pre-filter and then propelling it past a supply HEPA filter and toward the work zone to create a clean air environment. From there, a powerful DC ECM motor underneath the work zone will “pull” the air down through the air slots in the work surface to contain allergens. The air is then drawn past an exhaust pre-filter and pushed out an exhaust HEPA filter on the bottom of the unit. The Aeromax control system displays the performance of this “piston of air” by visualizing the plenum pressure as a series of multi-colored LED indicators. The NU-620 is designed as a dual-sided cabinet to provide flexibility to vivariums that have little room to maneuver their equipment.
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.