AD-IA NCPAP It is used to supply ventilation treatment for premature, neonates, infants and children below 14 years old. The different concentrations of medical oxygen will be configured by adjusting the proportion of oxygen and air. An inside alarm system of differential pressure is equipped. When alarming, the safety valve of this type of NCPAP machine will start automatically, and produce the mixed gas continuously to ensure the safety of patient. Adjust oxygen concentration from 21% to 100% optionally. With all-in-one design, AD-IA is easy to assemble and save space.
Working Principle of AD-IA NCPAP
Different concentrations of medical oxygen are configured by adjusting the ratio of oxygen to the air. The suitable oxygen concentration ratio enters the breathing tube through the humidifier and then enters the human body to support the patient's normal circulation oxygen breathing. The built-in air oxygen mixer is equipped with a differential pressure alarm valve. If the pressure difference between air and oxygen exceeds a certain ratio, the safety valve will start automatically when the alarm is given. Continuous production of mixed gases to ensure patient safety.
Features of AD-IA NCPAP
Wide selection of oxygen concentration, concentration adjustment from 21% to 100%, all-in-one machine design, easy assembly, space saving.
Built-in miniature air compressor for easy emergency transport
Specification of AD-IA NCPAP
FIO2 - 21-100% (±3%)
CPAP - 3-10cm H2O
Flow - 1-10LPM
Noise - ≤55dB(A)
Power Source - AC 220V/110V, 50-60Hz
Gas source - Air/Oxygen@0.3-0.4MPa
Alarm - Gas supply pressure difference > 0.1MPa