Safe AQ Angel Blood Glucose Monitoring System is designed for easy, user-friendly, convenient operation and only needs a small volume of blood sample. The Safe AQ Angel Test Strips require no coding which saves time and avoids human error due to improper operations. The memory mode allows you to save up to 200 blood glucose test results and 10 blood glucose control solution test results.
Features (Technical advantages)
1) Anti multiple interferences reliable results
2) Less blood volume required only 0.6μL
Performance Characteristics
The Safe AQ Angel blood glucose monitoring system complies with the requirements of ISO 15197:2013 ( In vitro diagnostic test systems- Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus).
Specification
Blood Volume
0.6μL
Sample Type
Capillary Whole Bloodyenous Whole Blood
Calibration
Plasma Equivalent
Meter Storage/Transportation Condition
-20℃~55℃
Dimension
103×57×22(mm)
Power Source
3VDC, 2 AAA Alkaline Batteries
Construction
Hand-held
Measurement Units
mg/dL or mmol/L
Measurement Range
20~600 mg/dL or 1.1~33.3mmol/L
Shelf Life
10 years(estimated by test 7 times per day). During use, user should maintain the product refer to this user manual’s requirements.
User performance evaluation:
A study evaluating glucose values from fingertip capillary blood samples obtained by 100 lay persons showed the following results:
100 % within ±0,83 mmol/L (±15 mg/ dL) of the YSI values at glucose concentrations below 5,55 mmol/l (100 mg/ dL), and 100 % within ±15 % of the YSI values at glucose concentrations at or above 5,55 mmol/L (100 mg/ dL).